Repeatability of in vitro power profile measurements for multifocal contact lenses
- PMID: 25704460
- DOI: 10.1016/j.clae.2015.01.013
Repeatability of in vitro power profile measurements for multifocal contact lenses
Abstract
Purpose: To evaluate the repeatability of an optical device (NIMO TR1504, Lambda-X, Belgium) for measuring multifocal contact lens power profiles.
Methods: The NIMO TR1504 was used to measure power profiles 30 times for each of 10 different contact lenses from 4 major companies. All contact lenses were labelled as -3D for distance vision; half were for high addition and half for low addition. The optical zone in all measurements was set to 3-mm radius. For each lens, the median power profile and the residuals of the 30 measurements were calculated. The 95% confidence bands and two indices that summarize measurement errors were calculated: the repeatability limit and an index of repeatability heterogeneity, quantifying heterogeneity of measurement errors over the optical zone.
Results: The repeatability limit was good (from 0.04D to 0.12D), for all multifocal contact lenses. Variability of measurement errors of power profiles was quite homogeneous along the optical zone for all lenses, although for some lenses variability was slightly higher in the centre than peripherally.
Conclusions: The repeatability of measured power profiles obtained by the NIMO TR1504 is lower than 0.12D for the multifocal contact lenses.
Keywords: Contact lens power profile; Multifocal contact lens; Repeatability; Schlieren principle.
Copyright © 2015 British Contact Lens Association. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Reliability of power profiles measured on NIMO TR1504 (Lambda-X) and effects of lens decentration for single vision, bifocal and multifocal contact lenses.J Optom. 2016 Apr-Jun;9(2):126-36. doi: 10.1016/j.optom.2015.10.005. Epub 2015 Nov 14. J Optom. 2016. PMID: 26590009 Free PMC article.
-
Power profiles of single vision and multifocal soft contact lenses.Cont Lens Anterior Eye. 2015 Feb;38(1):2-14. doi: 10.1016/j.clae.2014.07.008. Epub 2014 Aug 16. Cont Lens Anterior Eye. 2015. PMID: 25139719
-
Power profiles in multifocal contact lenses with variable multifocal zone.Clin Exp Optom. 2018 Jan;101(1):57-63. doi: 10.1111/cxo.12575. Epub 2017 Jul 18. Clin Exp Optom. 2018. PMID: 28718954
-
[Refraction and accommodation].Nippon Ganka Gakkai Zasshi. 2007 Feb;111(2):77-82. Nippon Ganka Gakkai Zasshi. 2007. PMID: 17338323 Review. Japanese.
-
Multifocal intraocular lenses.Ophthalmol Clin North Am. 2006 Mar;19(1):89-105, vi. doi: 10.1016/j.ohc.2005.09.002. Ophthalmol Clin North Am. 2006. PMID: 16500531 Review.
Cited by
-
On the power profiles of contact lenses measured with NIMO TR1504.J Optom. 2017 Oct-Dec;10(4):265-266. doi: 10.1016/j.optom.2016.10.002. Epub 2017 Jan 12. J Optom. 2017. PMID: 28089205 Free PMC article. No abstract available.
-
Power Profiles of Commercial Multifocal Soft Contact Lenses.Optom Vis Sci. 2017 Feb;94(2):183-196. doi: 10.1097/OPX.0000000000000998. Optom Vis Sci. 2017. PMID: 27748699 Free PMC article.
-
Reliability of power profiles measured on NIMO TR1504 (Lambda-X) and effects of lens decentration for single vision, bifocal and multifocal contact lenses.J Optom. 2016 Apr-Jun;9(2):126-36. doi: 10.1016/j.optom.2015.10.005. Epub 2015 Nov 14. J Optom. 2016. PMID: 26590009 Free PMC article.
-
Fourier tools for the evaluation of refractive multifocal designs.Sci Rep. 2023 Dec 19;13(1):22585. doi: 10.1038/s41598-023-50172-7. Sci Rep. 2023. PMID: 38114735 Free PMC article.
-
In vitro optical quality of monofocal aspheric toric intraocular lenses: effect of cylindrical power.Int Ophthalmol. 2018 Jun;38(3):933-941. doi: 10.1007/s10792-017-0537-z. Epub 2017 Apr 25. Int Ophthalmol. 2018. PMID: 28444524
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical