Light-distortion analysis as a possible indicator of visual quality after refractive lens exchange with diffractive multifocal intraocular lenses
- PMID: 25708210
- DOI: 10.1016/j.jcrs.2014.07.033
Light-distortion analysis as a possible indicator of visual quality after refractive lens exchange with diffractive multifocal intraocular lenses
Abstract
Purpose: To study the perception of light distortion after refractive lens exchange (RLE) with diffractive multifocal intraocular lenses (IOLs).
Setting: Clínica Oftalmológica das Antas, Porto, Portugal.
Design: Retrospective comparative study.
Methods: Refractive lens exchange was performed with implantation of an AT Lisa 839M (trifocal) or 909MP (bifocal toric) IOL, the latter if corneal astigmatism was more than 0.75 diopter (D). The postoperative visual and refractive outcomes were evaluated. A prototype light-distortion analyzer was used to quantify the postoperative light-distortion indices. A control group of eyes in which a Tecnis ZCB00 1-piece monofocal IOL was implanted had the same examinations.
Results: A trifocal or bifocal toric IOL was implanted in 66 eyes. The control IOL was implanted in 18 eyes. All 3 groups obtained a significant improvement in uncorrected distance visual acuity (UDVA) (P < .001) and corrected distance visual acuity (CDVA) (P = .001). The mean uncorrected near visual acuity (UNVA) was 0.123 logMAR with the trifocal IOL and 0.130 logMAR with the bifocal toric IOL. The residual refractive cylinder was less than 1.00 D in 86.7% of cases with the toric IOL. The mean light-distortion index was significantly higher in the multifocal IOL groups than in the monofocal group (P < .001), although no correlation was found between the light-distortion index and CDVA.
Conclusions: The multifocal IOLs provided excellent UDVA and functional UNVA despite increased light-distortion indices. The light-distortion analyzer reliably quantified a subjective component of vision distinct from visual acuity; it may become a useful adjunct in the evaluation of visual quality obtained with multifocal IOLs.
Financial disclosure: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Comparison of postoperative visual performance between bifocal and trifocal intraocular Lens based on randomized controlled trails: a meta-analysis.BMC Ophthalmol. 2019 Mar 14;19(1):78. doi: 10.1186/s12886-019-1078-1. BMC Ophthalmol. 2019. PMID: 30871503 Free PMC article.
-
Comparison of visual and refractive outcomes after bilateral implantation of toric intraocular lenses with or without a multifocal component.J Cataract Refract Surg. 2015 Jan;41(1):73-83. doi: 10.1016/j.jcrs.2014.04.032. Epub 2014 Nov 11. J Cataract Refract Surg. 2015. PMID: 25466485
-
One-year outcomes with new-generation multifocal intraocular lenses.Ophthalmology. 2008 Sep;115(9):1508-16. doi: 10.1016/j.ophtha.2008.04.017. Epub 2008 Jun 5. Ophthalmology. 2008. PMID: 18538402 Clinical Trial.
-
Visual and optical outcomes of a diffractive multifocal toric intraocular lens.J Cataract Refract Surg. 2013 Jul;39(7):1029-35. doi: 10.1016/j.jcrs.2013.02.037. Epub 2013 May 13. J Cataract Refract Surg. 2013. PMID: 23680631
-
[Multifocal intraocular lenses--an assessment of current status].Klin Monbl Augenheilkd. 1993 Jul;203(1):19-33. doi: 10.1055/s-2008-1045645. Klin Monbl Augenheilkd. 1993. PMID: 8411889 Review. German.
Cited by
-
Trifocal versus Bifocal Diffractive Intraocular Lens Implantation after Cataract Surgery or Refractive Lens Exchange: a Meta-analysis.J Korean Med Sci. 2018 Sep 27;33(44):e275. doi: 10.3346/jkms.2018.33.e275. eCollection 2018 Oct 29. J Korean Med Sci. 2018. PMID: 30369857 Free PMC article.
-
Comparison of postoperative visual performance between bifocal and trifocal intraocular Lens based on randomized controlled trails: a meta-analysis.BMC Ophthalmol. 2019 Mar 14;19(1):78. doi: 10.1186/s12886-019-1078-1. BMC Ophthalmol. 2019. PMID: 30871503 Free PMC article.
-
Through-Focus Vision Performance and Light Disturbances of 3 New Intraocular Lenses for Presbyopia Correction.J Ophthalmol. 2018 Jan 31;2018:6165493. doi: 10.1155/2018/6165493. eCollection 2018. J Ophthalmol. 2018. PMID: 29651343 Free PMC article.
-
Screening of positive dysphotopsia before multifocal intraocular lens implantation.Indian J Ophthalmol. 2024 Feb 1;72(Suppl 2):S211-S217. doi: 10.4103/IJO.IJO_1038_23. Epub 2024 Jan 25. Indian J Ophthalmol. 2024. PMID: 38271416 Free PMC article.
-
The Predictability of Preoperative Pilocarpine-Induced Lens Shift on the Outcomes of Accommodating Intraocular Lenses Implanted in Senile Cataract Patients.J Ophthalmol. 2016;2016:6127130. doi: 10.1155/2016/6127130. Epub 2016 Jul 19. J Ophthalmol. 2016. PMID: 27516899 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical