Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Clinical Trial
. 2015 Jun;135(6):549-56.
doi: 10.3109/00016489.2014.999874. Epub 2015 Feb 26.

Clinical applications of correlational vestibular autorotation test

Affiliations
Clinical Trial

Clinical applications of correlational vestibular autorotation test

Li-Chun Hsieh et al. Acta Otolaryngol. 2015 Jun.

Abstract

Conclusion: The correlational vestibular autorotation test (VAT) system has the advantages of good test-retest reliability and calibrations of absolute degrees of eye movement are unnecessary when acquiring a cross correlation coefficient (CCC). The approach is able to efficiently detect peripheral vestibulopathies.

Objective: A VAT has some drawbacks including poor test-retest reliability and slippage of sensor. This study aimed to develop a correlational VAT system and to evaluate the reliability and applicability of this system.

Methods: Twenty healthy participants and 10 vertiginous patients were enrolled. Vertical and horizontal autorotations from 0 to 3 Hz with either closed or open eyes were performed. A small sensor and a wireless transmission technique were used to acquire the electro-ocular graph and head velocity signals. The two signals were analyzed using CCCs to assess the functioning of the vestibular ocular reflex (VOR).

Results: The results showed a significantly greater CCC for open-eye versus closed-eye of head autorotations. The CCCs also increased significantly with head rotational frequencies. Moreover, the CCCs significantly correlated with the VOR gains at autorotation frequencies ≥1.0 Hz. The test-retest reliability was good (intraclass correlation coefficients ≥0.85). The vertiginous participants had significantly lower individual CCCs and overall average CCC than age- and-gender matched controls.

Keywords: Cross correlational analysis; electro-oculography; gyrometry; intraclass correlation.

PubMed Disclaimer

Publication types

LinkOut - more resources