. 2015 Jul;30(7):1265-8.
doi: 10.1016/j.arth.2015.02.019. Epub 2015 Feb 19.

A High Prevalence of Corrosion at the Head-Neck Taper with Contemporary Zimmer Non-Cemented Femoral Hip Components

Affiliations

A High Prevalence of Corrosion at the Head-Neck Taper with Contemporary Zimmer Non-Cemented Femoral Hip Components

Brian J McGrory et al. J Arthroplasty. 2015 Jul.

Abstract

Mechanically assisted crevice corrosion (MACC) occurs at metal/metal modular junctions in which at least one of the components is fabricated from cobalt-chromium alloy and may lead to adverse local tissue reaction (ALTR) in patients with metal-on-polyethylene (MoP) total hip arthroplasty. This type of reaction has been previously described in hips with head/neck modularity, but the prevalence is unknown. We found a prevalence of 1.1 percent in a consecutive series of 1356 contemporary Zimmer non-cemented femoral hip components followed for a minimum of 2years. The average time to presentation was 3.7years (range, 9-105months); delay in treatment led to irreversible soft tissue damage in three patients. We recommend usage of ceramic heads until this problem is further understood.

Keywords: THA; adverse local tissue reaction; failure; hip; metal-on-polymer.