Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography
- PMID: 25744093
- DOI: 10.1007/s00056-014-0278-9
Artifacts in orthodontic bracket systems in cone-beam computed tomography and multislice computed tomography
Abstract
Objectives: The purpose of this study was to quantify artifacts caused by different bracket systems in cone-beam computed tomography (CBCT) and multislice computed tomography (MSCT) scans.
Methods: Orthodontic brackets of four different systems were consecutively bonded to the surface of a residual molar on a human cadaveric mandible. One MSCT system and three CBCT units were used to scan each of the four bonded brackets, in addition to obtaining a blank reference scan of the tooth surface. All datasets were registered to the reference dataset using visualization software (Analyze 11.0® by AnalyzeDirect). Artifact-related reductions in image quality were expressed in percent of theoretical maximum standard deviations (SD) obtained for the gray values of the adjacent voxels, with higher percentages correlating more pronounced artifacts.
Results: Both the SD percentages for three defined line profiles and their mean values were almost invariably higher with the MSCT system than with the CBCT units. Looking into the individual SD percentages, two of the CBCT units (Pax Zenith 3D® and Picasso Trio®; both Vatech) produced higher values than the MSCT system (SOMATOM Definition AS+®; Siemens) in some line profiles. The titanium bracket, in particular, was associated with marked differences between the two scanner technologies, as the mean artifact intensities from this bracket were particularly high with the MSCT unit and relatively low with the CBCT units. The artifact intensities observed with the other three bracket systems varied widely depending on which scanner was used.
Conclusion: Different artifact intensities were noted depending on the composition of the bracket system and on the scanner technology (MSCT/CBCT). While the artifacts manifested themselves differently with different scanners, their adverse effects were comparable. However, given the variable severity of the artifacts observed depending on the materials scanned and the scanners used, a blanket recommendation for or against MSCT or CBCT units cannot be given on the basis of this study.
Similar articles
-
Cone beam computed tomography and low-dose multislice computed tomography in orthodontics and dentistry: a comparative evaluation on image quality and radiation exposure.J Orofac Orthop. 2014 Sep;75(5):384-98. doi: 10.1007/s00056-014-0232-x. Epub 2014 Aug 28. J Orofac Orthop. 2014. PMID: 25158951
-
Bone quality evaluation at dental implant site using multislice CT, micro-CT, and cone beam CT.Clin Oral Implants Res. 2015;26(1):e1-7. doi: 10.1111/clr.12315. Epub 2013 Dec 11. Clin Oral Implants Res. 2015. PMID: 24325572
-
Internal or in-scan validation: a method to assess CBCT and MSCT gray scales using a human cadaver.Oral Surg Oral Med Oral Pathol Oral Radiol. 2014 Jun;117(6):768-79. doi: 10.1016/j.oooo.2014.02.012. Epub 2014 Feb 24. Oral Surg Oral Med Oral Pathol Oral Radiol. 2014. PMID: 24842449
-
Dentomaxillofacial CBCT: Clinical Challenges for Indication-oriented Imaging.Semin Musculoskelet Radiol. 2020 Oct;24(5):479-487. doi: 10.1055/s-0040-1709428. Epub 2020 Oct 9. Semin Musculoskelet Radiol. 2020. PMID: 33036036 Review.
-
Technical aspects of dental CBCT: state of the art.Dentomaxillofac Radiol. 2015;44(1):20140224. doi: 10.1259/dmfr.20140224. Dentomaxillofac Radiol. 2015. PMID: 25263643 Free PMC article. Review.
Cited by
-
Robust and Accurate Mandible Segmentation on Dental CBCT Scans Affected by Metal Artifacts Using a Prior Shape Model.J Pers Med. 2021 May 1;11(5):364. doi: 10.3390/jpm11050364. J Pers Med. 2021. PMID: 34062762 Free PMC article.
-
Automatic Segmentation of Mandible from Conventional Methods to Deep Learning-A Review.J Pers Med. 2021 Jul 1;11(7):629. doi: 10.3390/jpm11070629. J Pers Med. 2021. PMID: 34357096 Free PMC article. Review.
-
Quantitative analysis of zirconia and titanium implant artefacts in three-dimensional virtual models of multi-slice CT and cone beam CT: does scan protocol matter?Dentomaxillofac Radiol. 2023 Nov;52(8):20230275. doi: 10.1259/dmfr.20230275. Epub 2023 Oct 23. Dentomaxillofac Radiol. 2023. PMID: 37641962 Free PMC article.
-
Fusion of intra-oral scans in cone-beam computed tomography scans.Clin Oral Investig. 2021 Jan;25(1):77-85. doi: 10.1007/s00784-020-03336-y. Epub 2020 Jun 3. Clin Oral Investig. 2021. PMID: 32495223 Free PMC article.
-
Comparison of feasibility, time consumption and costs of three virtual planning systems for surgical correction of midfacial deficiency.Maxillofac Plast Reconstr Surg. 2021 Jan 7;43(1):2. doi: 10.1186/s40902-020-00284-1. Maxillofac Plast Reconstr Surg. 2021. PMID: 33411020 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials