Benchmark IMRT evaluation of a Co-60 MRI-guided radiation therapy system
- PMID: 25746353
- DOI: 10.1016/j.radonc.2015.01.015
Benchmark IMRT evaluation of a Co-60 MRI-guided radiation therapy system
Abstract
A device for MRI-guided radiation therapy (MR-IGRT) that uses cobalt-60 sources to deliver intensity modulated radiation therapy is now commercially available. We investigated the performance of the treatment planning and delivery system against the benchmark recommended by the American Association of Physicists in Medicine (AAPM) Task Group 119 for IMRT commissioning and demonstrated that the device plans and delivers IMRT treatments within recommended confidence limits and with similar accuracy as linac IMRT.
Keywords: Co-60; IMRT; MRI-Guided radiation therapy.
Copyright © 2015 Elsevier Ireland Ltd. All rights reserved.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
