Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2015 Jun;12(6):1169-76.
doi: 10.1016/j.hrthm.2015.02.035. Epub 2015 Mar 4.

The clinical course of patients with implantable cardioverter-defibrillators: Extended experience on clinical outcome, device replacements, and device-related complications

Affiliations

The clinical course of patients with implantable cardioverter-defibrillators: Extended experience on clinical outcome, device replacements, and device-related complications

Aafke C van der Heijden et al. Heart Rhythm. 2015 Jun.

Abstract

Background: Large randomized trials demonstrated the beneficial effect of implantable cardioverter-defibrillator (ICD) and cardiac resynchronization therapy-defibrillator (CRT-D) treatments in selected patients. Data on long-term follow-up of patients outside the setting of clinical trials are scarce.

Objective: The aim of this study was to evaluate the long-term outcome of ICD and CRT-D recipients.

Methods: All patients who underwent ICD (n = 1729 [57%]) or CRT-D (n = 1326 [43%]) implantation at the Leiden University Medical Center since 1996 were evaluated. Follow-up visits were performed every 3-6 months, and events were registered. Cumulative incidence curves of device therapy and device-related complications were adjusted for the competing risk of all-cause mortality.

Results: After a median follow-up of 5.1 years (25th-75th percentile 3.1-7.8 years), 842 patients (28%) died. The cumulative incidence of all-cause mortality was 49% (95% confidence interval [CI] 45%-54%) in ICD recipients after 12 years of follow-up and 55% (95% CI 52%-58%) in CRT-D recipients after 8 years of follow-up. A total of 1081 patients (35%) received appropriate defibrillator therapy. The cumulative incidence of appropriate therapy in ICD patients was 58% (95% CI 54%-62%) after 12 years of follow-up and 39% (95% CI 35%-43%) in CRT-D patients after 8 years of follow-up. Twelve-year cumulative incidences of adverse events were 20% (95% CI 18%-22%) for inappropriate shock, 6% (95% CI 5%-8%) for device-related infection, and 17% (95% CI 14%-21%) for lead failure.

Conclusion: After long-term follow-up of ICD (12 years) and CRT-D (8 years) recipients, 49% of ICD recipients and 55% of CRT-D recipients had died. Appropriate ICD therapy was received by the majority (58%) of ICD recipients and by almost 40% of CRT-D recipients.

Keywords: Cardiac resynchronization therapy; Defibrillator therapy; Device replacements; Device-related complications; Implantable cardioverter-defibrillator.