. 1988 Nov;2(11):249-56.
doi: 10.1002/rcm.1290021111.

The determination of protein, oligonucleotide and peptide molecular weights by ion-spray mass spectrometry

Affiliations

The determination of protein, oligonucleotide and peptide molecular weights by ion-spray mass spectrometry

T R Covey et al. Rapid Commun Mass Spectrom. 1988 Nov.

Abstract

The mass spectra of several compounds with molecular weights in the 2500-20,000 Da range were obtained with a quadrupole mass spectrometer equipped with an atmospheric pressure ion source. Average molecular weight determinations of mellitin (2846.4 Da), a synthetic oligonucleotide (4262.8 Da), myoglobin (16,950.4 Da) and on the subunits of beta-lactoglobulin (18,277.1 Da) requiring as little as 1 pmol of material were achieved with accuracies and precisions of +/- 1 Da. An ion-spray interface was used to produce ions via the ion evaporation process, producing mass spectra containing a series of multiply-charged molecular species. A simple method for calculating the molecular weight of unknown compounds from the spectra containing multiply-charged ions is described.