Evaluation of the use of automatic exposure control and automatic tube potential selection in low-dose cerebrospinal fluid shunt head CT
- PMID: 25779098
- DOI: 10.1007/s00234-015-1508-6
Evaluation of the use of automatic exposure control and automatic tube potential selection in low-dose cerebrospinal fluid shunt head CT
Abstract
Introduction: Cerebrospinal fluid shunts are primarily used for the treatment of hydrocephalus. Shunt complications may necessitate multiple non-contrast head CT scans resulting in potentially high levels of radiation dose starting at an early age. A new head CT protocol using automatic exposure control and automated tube potential selection has been implemented at our institution to reduce radiation exposure. The purpose of this study was to evaluate the reduction in radiation dose achieved by this protocol compared with a protocol with fixed parameters.
Methods: A retrospective sample of 60 non-contrast head CT scans assessing for cerebrospinal fluid shunt malfunction was identified, 30 of which were performed with each protocol. The radiation doses of the two protocols were compared using the volume CT dose index and dose length product. The diagnostic acceptability and quality of each scan were evaluated by three independent readers.
Results: The new protocol lowered the average volume CT dose index from 15.2 to 9.2 mGy representing a 39 % reduction (P < 0.01; 95 % CI 35-44 %) and lowered the dose length product from 259.5 to 151.2 mGy/cm representing a 42 % reduction (P < 0.01; 95 % CI 34-50 %). The new protocol produced diagnostically acceptable scans with comparable image quality to the fixed parameter protocol.
Conclusion: A pediatric shunt non-contrast head CT protocol using automatic exposure control and automated tube potential selection reduced patient radiation dose compared with a fixed parameter protocol while producing diagnostic images of comparable quality.
Similar articles
-
The Role of Limited Head Computed Tomography in the Evaluation of Pediatric Ventriculoperitoneal Shunt Malfunction.Pediatr Emerg Care. 2016 Sep;32(9):585-9. doi: 10.1097/PEC.0000000000000760. Pediatr Emerg Care. 2016. PMID: 27299297
-
Low-dose protocol for head CT in monitoring hydrocephalus in children.Med Sci Monit. 2007 May;13 Suppl 1:147-51. Med Sci Monit. 2007. PMID: 17507900
-
Reducing the radiation exposure from CT scanning in children with shunts: a nationwide survey and a departmental CT protocol.Br J Neurosurg. 2018 Oct;32(5):558-562. doi: 10.1080/02688697.2018.1485874. Epub 2018 Jul 18. Br J Neurosurg. 2018. PMID: 30019596
-
Radiographic evaluation of pediatric cerebrospinal fluid shunt malfunction in the emergency setting.Pediatr Emerg Care. 2015 Jun;31(6):435-40; quiz 441-3. doi: 10.1097/PEC.0000000000000462. Pediatr Emerg Care. 2015. PMID: 26035499 Review.
-
CT scan exposure in children with ventriculo-peritoneal shunts: single centre experience and review of the literature.Childs Nerv Syst. 2020 Mar;36(3):591-599. doi: 10.1007/s00381-019-04345-3. Epub 2019 Aug 19. Childs Nerv Syst. 2020. PMID: 31428872 Review.
Cited by
-
Radiation dose reduction using 100-kVp and a sinogram-affirmed iterative reconstruction algorithm in adolescent head CT: Impact on grey-white matter contrast and image noise.Eur Radiol. 2017 Jul;27(7):2717-2725. doi: 10.1007/s00330-016-4679-6. Epub 2016 Dec 13. Eur Radiol. 2017. PMID: 27966043
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical