Review
. 2015 Jun;9(3):95-102.
doi: 10.1177/1753944715575970. Epub 2015 Mar 16.

A balloon-expandable sheath facilitates transfemoral TAVR in patients with peripheral vascular disease and tortuosity

Affiliations
Free article
Review

A balloon-expandable sheath facilitates transfemoral TAVR in patients with peripheral vascular disease and tortuosity

Andrew M Goldsweig et al. Ther Adv Cardiovasc Dis. 2015 Jun.
Free article

Abstract

Objectives: We sought to perform transcatheter aortic valve replacement (TAVR) via the transfemoral approach in patients with peripheral arterial disease (PAD), small caliber ileofemoral vessels and vascular tortuosity.

Background: For patients with increased surgical risk, TAVR is associated with a higher 1-year survival rate than surgical aortic valve replacement (SAVR). Transfemoral vascular access for TAVR results in superior outcomes versus procedures performed via other routes in terms of mortality, morbidity and healthcare economics. In many patients, the ability to safely perform the procedure via the transfemoral approach is limited by narrow, diseased and tortuous ileofemoral vasculature.

Methods: We employed the SoloPath Balloon Expandable TransFemoral Access System (Terumo Med. Corp., Tokyo, Japan) to perform transfemoral TAVR in five patients with PAD, small caliber ileofemoral vessels and vascular tortuosity.

Results: We report our experience using this balloon-expandable sheath during 5 cases of transfemoral TAVR in patients with inhospitable ileofemoral vasculature of mean diameter ⩽ 5.8 mm. The unexpanded sheath's malleable structure and hydrophilic coating permitted deployment despite severe stenoses and tortuosity. Subsequent inflation to 18 Fr facilitated successful TAVR. Postprocedural angiography demonstrated no significant vascular access complications. In one case, the entire procedure was performed percutaneously, without common femoral artery surgical cutdown.

Conclusions: The SoloPath sheath system permits transfemoral TAVR in patients with PAD small caliber ileofemoral vessels and vascular tortuosity. The transfemoral balloon-expandable sheath allowed these patients to avoid the increased morbidity and mortality risks associated with direct aortic or transapical access.

Keywords: PAD; SoloPath; TAVR; balloon-expandable; transfemoral.