. 2016 Jun;23(3):425-35.
doi: 10.1007/s12350-015-0097-4. Epub 2015 Mar 19.

Quantitative iodine-123-metaiodobenzylguanidine (MIBG) SPECT imaging in heart failure with left ventricular systolic dysfunction: Development and validation of automated procedures in conjunction with technetium-99m tetrofosmin myocardial perfusion SPECT

Affiliations

Quantitative iodine-123-metaiodobenzylguanidine (MIBG) SPECT imaging in heart failure with left ventricular systolic dysfunction: Development and validation of automated procedures in conjunction with technetium-99m tetrofosmin myocardial perfusion SPECT

Ian P Clements et al. J Nucl Cardiol. 2016 Jun.

Abstract

Background: The purpose of this study was to develop and validate new approaches to quantitative MIBG myocardial SPECT imaging in heart failure (HF) subjects.

Methods and results: Quantitative MIBG myocardial SPECT analysis methods, alone and in conjunction with 99mTc-tetrofosmin perfusion SPECT, were adapted from previously validated techniques for the analysis of SPECT and PET perfusion imaging. To account for underestimation of MIBG defect severity in subjects with global reduction in uptake, a mixed reference database based on planar heart/mediastinum (H/M) ratio categories was used. Extent and severity of voxel-based defects and number of myocardial segments with significant dysinnervation (derived score ≥2) were determined. MIBG/99mTc-tetrofosmin mismatch was quantified using regions with preserved innervation as the reference for scaling 99mTc-tetrofosmin voxel maps. Quantification techniques were tested on studies of 619 ischemic (I) and 319 non-ischemic (NI) HF subjects. Using all analytical techniques, IHF subjects had significantly greater and more severe MIBG SPECT abnormalities compared with NIHF subjects. Innervation/perfusion mismatches were also larger in IHF subjects. Findings were consistent between voxel- and myocardial-segment-based quantitation methods.

Conclusions: Multiple objective methods for quantitation of MIBG SPECT imaging studies provided internally consistent results for distinguishing the different patterns of uptake between IHF and NIHF subjects.

Keywords: MIBG imaging; SPECT; heart failure.

PubMed Disclaimer

Comment in

  • It's not all in the numbers.
    Travin MI. Travin MI. J Nucl Cardiol. 2016 Jun;23(3):436-41. doi: 10.1007/s12350-015-0105-8. Epub 2015 Mar 24. J Nucl Cardiol. 2016. PMID: 25802176 No abstract available.

References

    1. Am Heart J. 1993 Aug;126(2):390-8 - PubMed
    1. J Nucl Cardiol. 2003 Mar-Apr;10 (2):121-31 - PubMed
    1. J Nucl Med. 1999 Jun;40(6):917-23 - PubMed
    1. Circulation. 1990 Oct;82(4):1095-105 - PubMed
    1. JACC Cardiovasc Imaging. 2013 Dec;6(12):1327-41 - PubMed

MeSH terms

Substances

LinkOut - more resources