Randomized comparison of the i-gel(TM) with the LMA Supreme (TM) in anesthetized adult patients
- PMID: 25801488
- DOI: 10.1007/s00101-015-0020-z
Randomized comparison of the i-gel(TM) with the LMA Supreme (TM) in anesthetized adult patients
Abstract
Background: The LMA Supreme(TM) (LMA-S) and i-gel(TM) are two of the most commonly used supraglottic airway devices (SADs) with an inbuilt drain channel. These devices are particularly indicated for performing certain procedures accompanied by high peak airway pressure, such as laparoscopy. This study compared the devices regarding efficacy, safety, ease of use and incidence of adverse events, focusing on the postoperative rate of sore throat, dysphagia or dysphonia and development with time, in patients undergoing laparoscopic cholecystectomy procedures under general anesthesia.
Methods: This was a prospective, randomized, controlled clinical study including 140 patients randomized into 2 groups undergoing elective laparoscopic cholecystectomy to use either i-gel or LMA-S. After the general anesthesia procedure, the speed of insertion, success rates, ease of insertion of the drain tube, leak pressure and tidal volume achieved by the devices were evaluated. The postoperative oropharyngeal discomfort (POPD) during the period of stay of the patients in the recovery room was also recorded.
Results: The mean leak pressure was comparable between the two groups (i-gel 28.18 ± 3.90 cmH2O and LMA-S 27.50 ± 4 cmH2O, p = 0.09), as well as maximum expiratory tidal volume provided (i-gel 559.60 ± 45.25 ml and LMA-S 548.95 ± 56.18 ml, p = 0.12). Insertion times were lower for the i-gel (10 ± 1.62 s) compared with the LMA-S (11.31 ± 2.85 s, p = 0.008). Insertion success rate at the first attempt was higher for the LMA-S (95 % compared with i-gel 79 %, p = 0.007). Drain tubes were easier to insert in the LMA-S group (p < 0.001). No differences were found between groups relating to intraoperative complications. Frequency of coughing and visible blood on removal of the device were low and comparable in both groups (p = 0.860 and p = 0.623, respectively). There were no differences relating to the incidence of sore throat, dysphagia or hoarseness at 10 min postoperatively between groups (p = 0.088). The i-gel group complained about a higher sore throat score at 2 h postoperatively (p = 0.009), specifically patients receiving i-gel suffered more from sore throats with 0.24 more points on the visual analog scale (VAS) than patients from the LMA-S group. The i-gel group also reported a lower POPD drop during the first 2 h (p < 0.001).
Conclusion: No differences were found between i-gel and LMA-S regarding leak pressure in the groups of anesthetized patients undergoing laparoscopic cholecystectomy. The LMA-S was easier to insert than the i-gel (based on its better first time success rate) and this device showed better ease of drain tube insertion, although the i-gel was quicker to insert than the LMA-S. The i-gel resulted in higher sore throat scores at 2 h postoperatively and lower POPD reduction during the 2 h period studied in the recovery room was reported.
Similar articles
-
Use of second generation supra-glottic airway devices during laparoscopic cholecystectomy: a prospective, randomized comparison of LMA Proseal™, LMA SupremeTM and igel™.Acta Anaesthesiol Belg. 2016;67(3):121-128. Acta Anaesthesiol Belg. 2016. PMID: 29873467 Clinical Trial.
-
Sore throat following three adult supraglottic airway devices: A randomised controlled trial.Eur J Anaesthesiol. 2017 Jul;34(7):417-424. doi: 10.1097/EJA.0000000000000539. Eur J Anaesthesiol. 2017. PMID: 27755181 Clinical Trial.
-
Comparison of Laryngeal Mask Airway Supreme and Laryngeal Mask Airway Proseal with respect to oropharyngeal leak pressure during laparoscopic cholecystectomy: a randomised controlled trial.Eur J Anaesthesiol. 2013 Mar;30(3):119-23. doi: 10.1097/EJA.0b013e32835aba6a. Eur J Anaesthesiol. 2013. PMID: 23318811 Clinical Trial.
-
Laryngeal mask airway ProSeal provides higher oropharyngeal leak pressure than i-gel in adult patients under general anesthesia: a meta-analysis.J Clin Anesth. 2016 Sep;33:298-305. doi: 10.1016/j.jclinane.2016.04.020. Epub 2016 May 18. J Clin Anesth. 2016. PMID: 27555181 Review.
-
Role of laryngeal mask airway in laparoscopic cholecystectomy.World J Gastrointest Surg. 2015 Nov 27;7(11):319-25. doi: 10.4240/wjgs.v7.i11.319. World J Gastrointest Surg. 2015. PMID: 26649155 Free PMC article. Review.
Cited by
-
Comparison of oropharyngeal leak pressure of air-Q™, i-gel™, and laryngeal mask airway supreme™ in adult patients during general anesthesia: A randomized controlled trial.Saudi J Anaesth. 2017 Oct-Dec;11(4):390-395. doi: 10.4103/sja.SJA_149_17. Saudi J Anaesth. 2017. PMID: 29033717 Free PMC article.
-
[Update Mallampati : Theoretical and practical knowledge of European anesthetists on basic evaluation of airways].Anaesthesist. 2018 Oct;67(10):738-744. doi: 10.1007/s00101-018-0481-y. Epub 2018 Aug 31. Anaesthesist. 2018. PMID: 30171286 German.
-
An audit of post-operative sore throat using different laryngeal mask airways.Indian J Anaesth. 2020 Jun;64(6):513-516. doi: 10.4103/ija.IJA_963_19. Epub 2020 Jun 1. Indian J Anaesth. 2020. PMID: 32792717 Free PMC article. No abstract available.
-
Thyromental distance ("Patil") revisited : Knowledge and performance of a basic airway screening tool among European anesthetists.Anaesthesist. 2018 Mar;67(3):198-203. doi: 10.1007/s00101-018-0412-y. Epub 2018 Feb 1. Anaesthesist. 2018. PMID: 29392357 English.