Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Observational Study
. 2016 Feb;36(1):9-15.
doi: 10.1007/s10792-015-0067-5. Epub 2015 Mar 28.

Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication

Affiliations
Observational Study

Agreement between diurnal variations of intraocular pressure by Tono-Pen and Goldmann applanation tonometer in patients on topical anti-glaucoma medication

Shikha Gupta et al. Int Ophthalmol. 2016 Feb.

Abstract

To estimate agreement in diurnal variations of intraocular pressure (IOP) by Tono-Pen (TP) and Goldmann applanation tonometer (GAT) in glaucoma patients on topical anti-glaucoma medication(s). IOP was measured at every 3 h from 7 a.m. to 10 a.m. in 50 eyes of glaucoma patients on topical medication(s). Diurnal fluctuation of IOP by each method was calculated as maximum-minimum IOP in a day. Central corneal thickness (CCT) was measured by ultrasonic pachymeter. There was good correlation between TP and GAT at all times during a day, minimum, and maximum IOPs during a day (Correlation coefficient, 0.706 at 7 a.m., 0.624 at 10 a.m., 0.682 at 1 p.m., 0.814 at 4 p.m., 0.652 at 7 p.m., 0.572 at 10 p.m., 0.668 minimum IOP, 0.689 maximum IOP). Mean IOPs by TP were always higher than GAT at all times during a day. Bland-Altman plots suggested a close relationship between the two sets of readings, and that this relationship was consistent at different times in a day, in maximum IOPs, minimum IOPs and also in fluctuation of IOPs. Linear regression analysis between the differences of diurnal fluctuation (diurnal fluctuation by GAT-diurnal fluctuation by TP) and CCT showed strong association (R 2 = 0.857, p < 0.001). The mean change in difference of diurnal fluctuation (GAT-TP) for a 10-micron increase in CCT was 0.69 mmHg. TP can be considered a reliable alternative to GAT in glaucoma patients for knowing the diurnal control of IOP; however these two methods should not be used interchangeably. Difference of diurnal fluctuation between two methods is dependent on CCT.

Keywords: Central corneal thickness; Diurnal fluctuation; Glaucoma; Goldmann applanation; Intraocular pressure; Tono-Pen.

PubMed Disclaimer

References

    1. Arch Ophthalmol. 1963 Sep;70:302-11 - PubMed
    1. Acta Ophthalmol Scand. 2007 May;85(3):272-9 - PubMed
    1. Am J Ophthalmol. 1998 Oct;126(4):487-97 - PubMed
    1. Surv Ophthalmol. 1993 Jul-Aug;38(1):1-30 - PubMed
    1. Klin Monbl Augenheilkd. 2002 Nov;219(11):785-90 - PubMed