Interactive near-real-time high-resolution imaging for MR-guided lumbar interventions using ZOOM imaging in an open 1.0 Tesla MRI system--initial experience
- PMID: 25830904
- DOI: 10.1515/bmt-2014-0118
Interactive near-real-time high-resolution imaging for MR-guided lumbar interventions using ZOOM imaging in an open 1.0 Tesla MRI system--initial experience
Abstract
Purpose: Different techniques for magnetic resonance-guided lumbar interventions have been introduced in recent years. Appropriate pulse sequence design is crucial since high spatial resolution often comes at the cost of lower temporal resolution. The purpose of this study was to evaluate the value of accelerated reduced field of view (ZOOM)-based imaging sequences for lumbar interventions.
Methods: ZOOM imaging was used in 31 interventions (periradicular, facet joint, epidural infiltrations, and discography) performed in 24 patients (10 women, 14 men; age 43 ± 13.3 years). Signal-to-noise ratio and contrast-to-noise ratio (CNR) were determined and retrospectively compared with standard preinterventional (T2 weighted), peri-interventional (proton density), and postinterventional (spectral presaturation with inversion recovery [SPIR]) imaging. Needle artifacts were assessed by direct measurement as well as with parallel and perpendicular needle profiles. Puncture times were compared to similar interventions previously performed in our department.
Results: No significant differences in signal intensities (standard/ZOOM: 152.0/151.6; p=0.136) and CNR values (2.0/4.0; p=0.487) were identified for T2-weighted sequences. The needle artifact signal intensity was comparable (648.1/747.5; p=0.172) for peri-interventional imaging. Standard interventional (fat needle: 43.8/23.4; p<0.001; muscle needle: 6.2/2.4; p<0.001) and SPIR sequences (43.3/13.9; p=0.010) showed a higher CNR than corresponding ZOOM sequences did. Needle artifacts were larger in ZOOM (2.4 mm/2.9 mm; p=0.005). The profiles revealed that ZOOM imaging delivers more overall signal intensity. The turning points of both profiles were comparable. ZOOM reduced intervention times significantly (329.1 s/228.5 s; p=0.026).
Conclusion: ZOOM imaging is a feasible interactive sequence for lumbar interventions. It ameliorates the tradeoff between image quality and temporal resolution. Moreover, the sequence design reduces intervention times significantly.
Similar articles
-
Initial clinical experience with a quadrupole butterfly coil for spinal injection interventions in an open MRI system at 1.0 tesla.Biomed Tech (Berl). 2014 Feb;59(1):39-45. doi: 10.1515/bmt-2013-0066. Biomed Tech (Berl). 2014. PMID: 24152888 Clinical Trial.
-
Evaluation of magnetic resonance imaging-compatible needles and interactive sequences for musculoskeletal interventions using an open high-field magnetic resonance imaging scanner.Cardiovasc Intervent Radiol. 2010 Apr;33(2):346-51. doi: 10.1007/s00270-009-9676-6. Epub 2009 Aug 25. Cardiovasc Intervent Radiol. 2010. PMID: 19705200
-
Percutaneous abscess drainage using near real-time MR guidance in an open 1.0-T MR scanner: proof of concept.Invest Radiol. 2013 Jun;48(6):477-84. doi: 10.1097/RLI.0b013e318284383f. Invest Radiol. 2013. PMID: 23462676 Clinical Trial.
-
[MR-Guided pain therapy: principles and clinical applications].Rofo. 2007 Sep;179(9):914-24. doi: 10.1055/s-2007-963200. Rofo. 2007. PMID: 17705114 Review. German.
-
Magnetic resonance imaging-guided spine injections.Top Magn Reson Imaging. 2011 Aug;22(4):143-51. doi: 10.1097/RMR.0b013e31827e5de1. Top Magn Reson Imaging. 2011. PMID: 23514922 Review.
Cited by
-
MR-guided lumbar facet radiofrequency denervation for treatment of patients with chronic low back pain in an open 1.0 Tesla MRI system.Ann Transl Med. 2021 Jul;9(13):1056. doi: 10.21037/atm-21-633. Ann Transl Med. 2021. PMID: 34422968 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources