Dynamic 31P MR spectroscopy of plantar flexion: influence of ergometer design, magnetic field strength (3 and 7 T), and RF-coil design
- PMID: 25832057
- DOI: 10.1118/1.4914448
Dynamic 31P MR spectroscopy of plantar flexion: influence of ergometer design, magnetic field strength (3 and 7 T), and RF-coil design
Abstract
Purpose: Dynamic phosphorus magnetic resonance spectroscopy ((31)P MRS) during and after acute exercise enables the noninvasive in vivo determination of the mitochondrial capacity of skeletal muscle. Nevertheless, the lack of standardization in experimental setups leads to significant variations in published values of maximal aerobic capacity, even in the population of healthy volunteers. Thus, in this study, we aimed to assess the impact of the ergometer type (pneumatic and mechanical resistance construction), radiofrequency (RF)-coil diameter, and different magnetic field strengths (3 and 7 T) on the metabolic parameters measured by dynamic (31)P MRS during a plantar flexion isotonic exercise protocol within the same group of healthy volunteers.
Methods: Dynamic (31)P MRS measurements of the calf muscle in 11 volunteers (mean age, 36 ± 13 yrs; mean BMI, 23.5 ± 2.5 kg/m(2)), on a 3 T MR system with a custom-made mechanical ergometer in the first research laboratory (RL1) and on 3 and 7 T MR systems equipped with a commercial pneumatic ergometer in the second research laboratory (RL2), were performed at three different workloads. RF-coils differed slightly between the sites and MR systems used. The repeatability of the experimental protocol was tested in every setup. The basal concentrations of phosphocreatine (PCr), exercise-induced depletion of PCr (ΔPCr), initial PCr resynthesis rate (VPCr), and mitochondrial capacity (Qmax) were calculated and compared between the research sites and field strengths.
Results: High repeatability of the measurement protocol was found in every experimental setup. No significant differences at any workload were found in these metabolic parameters assessed at different magnetic field strengths (3 T vs 7 T), using the same ergometer (in RL2) and a similar RF-coil. In the inter-research laboratory comparison at the same field strength (3 T), but with using different ergometers and RF-coils, differences were found in the concentration of PCr measured at rest and in the drop in PCr signal intensity. These differences translated into difference in the value of mitochondrial capacity at a workload of 15% of maximal voluntary contraction (MVC) force (0.45 ± 0.16 mM/s vs 0.31 ± 0.08 mM/s, in the RL1 and RL2, respectively).
Conclusions: Metabolic parameters measured during exercise challenge by dynamic (31)P MRS do not depend upon the magnetic field strength used. For multicenter studies with different ergometers, it is important to set the same workload, measurement, and evaluation protocols, especially when the effects of very mild exercise (15% MVC) are to be compared. However, a higher workload (24% MVC) decreases the influence of imperfections and intersite differences for the assessed value of maximal mitochondrial capacity.
Similar articles
-
MR-compatible pedal ergometer for reproducible exercising of the human calf muscle.Med Eng Phys. 2014 Jul;36(7):933-7. doi: 10.1016/j.medengphy.2014.02.026. Epub 2014 Apr 2. Med Eng Phys. 2014. PMID: 24703504
-
A low-cost Mr compatible ergometer to assess post-exercise phosphocreatine recovery kinetics.MAGMA. 2017 Jun;30(3):281-289. doi: 10.1007/s10334-016-0605-9. Epub 2017 Jan 4. MAGMA. 2017. PMID: 28054143 Free PMC article.
-
High-energy phosphate metabolism during incremental calf exercise in humans measured by 31 phosphorus magnetic resonance spectroscopy (31P MRS).Magn Reson Imaging. 2004 Jan;22(1):109-15. doi: 10.1016/j.mri.2003.07.001. Magn Reson Imaging. 2004. PMID: 14972400
-
Absolute quantification of phosphorus metabolite concentrations in human muscle in vivo by 31P MRS: a quantitative review.NMR Biomed. 2007 Oct;20(6):555-65. doi: 10.1002/nbm.1192. NMR Biomed. 2007. PMID: 17628042 Review.
-
MR compatible ergometers for dynamic 31P MRS.J Appl Biomed. 2019 Jun;17(2):91-98. doi: 10.32725/jab.2019.006. Epub 2019 Apr 15. J Appl Biomed. 2019. PMID: 34907736 Review.
Cited by
-
Effects of a 12-Week Interventional Exercise Programme on Muscle Strength, Mobility and Fitness in Patients With Diabetic Foot in Remission: Results From BIONEDIAN Randomised Controlled Trial.Front Endocrinol (Lausanne). 2022 Jul 5;13:869128. doi: 10.3389/fendo.2022.869128. eCollection 2022. Front Endocrinol (Lausanne). 2022. PMID: 35865313 Free PMC article. Clinical Trial.
-
1H and 31P MR Spectroscopy to Assess Muscle Mitochondrial Dysfunction in Long COVID.Radiology. 2024 Dec;313(3):e233173. doi: 10.1148/radiol.233173. Radiology. 2024. PMID: 39718498
-
Adiabatic excitation for 31 P MR spectroscopy in the human heart at 7 T: A feasibility study.Magn Reson Med. 2017 Nov;78(5):1667-1673. doi: 10.1002/mrm.26576. Epub 2016 Dec 21. Magn Reson Med. 2017. PMID: 28000961 Free PMC article.
-
Using a whole-body 31P birdcage transmit coil and 16-element receive array for human cardiac metabolic imaging at 7T.PLoS One. 2017 Oct 26;12(10):e0187153. doi: 10.1371/journal.pone.0187153. eCollection 2017. PLoS One. 2017. PMID: 29073228 Free PMC article.
-
Regionally resolved cardiac metabolism using a dipole-loop array coil for 7 T 31P-MRSI.Magn Reson Med. 2025 Aug;94(2):480-496. doi: 10.1002/mrm.30492. Epub 2025 Mar 23. Magn Reson Med. 2025. PMID: 40123193 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources