. 2015 Jun;84(6):1068-74.
doi: 10.1016/j.ejrad.2015.03.014. Epub 2015 Mar 23.

Ultra-low iodine concentrations iso-attenuating with diagnostic 0.5M gadolinium in endovascular procedures to minimize the risk of contrast nephropathy: A phantom study

Affiliations

Ultra-low iodine concentrations iso-attenuating with diagnostic 0.5M gadolinium in endovascular procedures to minimize the risk of contrast nephropathy: A phantom study

Jakob Swanberg et al. Eur J Radiol. 2015 Jun.

Abstract

Purpose: To establish the concentrations of iodine contrast media (I-CM) iso-attenuating with 0.5M gadolinium contrast media (Gd-CM), regarded diagnostic in catheter angiography and vascular interventions in azotemic patients, at various X-ray tube potentials with correlation to skin radiation dose.

Materials and method: 20-mL syringes filled with 30, 40, 50, 70 and 90 mgI/mL, 0.5M Gd-CM and air were placed in a water-equivalent phantom and exposed at about 50, 60, 70, 80 and 90 kV in an X-ray angiographic system. Relative contrast between the contrast materials and the background phantom material was measured on a PACS workstation. Radiation entrance dose, measured with a dose meter and estimated from the dose-area-product (DAP), was adjusted for radiation backscatter to simulate absorbed skin dose.

Result: The iodine concentrations 30, 40, 50, 70 and 90 mg/mL resulted in the same relative contrast as 0.5M gadolinium at 53, 57, 62, 71 and 85 kVp, respectively. Air had lower relative contrast than all iodine concentrations at all kVp-settings except for 30 mgI/mL above 84 kVp. The measured skin radiation dose was less than 1 mGy per exposure at all kVp-settings, and around 25-30% lower than the dose estimations derived from the angiographic system's in-built DAP meter.

Conclusion: Low-kilovoltage X-ray technique and ultra-low concentrations of I-CM iso-attenuating with 0.5M Gd-CM may be utilizable in peripheral arteriography and endovascular interventions, to minimize the total CM-dose to avoid CIN in azotemic patients.

Keywords: Angiography; Contrast media; Contrast nephropathy; Renal failure.