Clinical Assessment of Emission- and Segmentation-Based MR-Guided Attenuation Correction in Whole-Body Time-of-Flight PET/MR Imaging
- PMID: 25858043
- DOI: 10.2967/jnumed.115.154807
Clinical Assessment of Emission- and Segmentation-Based MR-Guided Attenuation Correction in Whole-Body Time-of-Flight PET/MR Imaging
Abstract
The joint maximum-likelihood reconstruction of activity and attenuation (MLAA) for emission-based attenuation correction has regained attention since the advent of time-of-flight PET/MR imaging. Recently, we improved the performance of the MLAA algorithm using an MR imaging-constrained gaussian mixture model (GMM). In this study, we compare the performance of our proposed algorithm with standard 4-class MR-based attenuation correction (MRAC) implemented on commercial systems.
Methods: Five head and neck (18)F-FDG patients were scanned on PET/MR imaging and PET/CT scanners. Dixon fat and water MR images were registered to CT images. MRAC maps were derived by segmenting the MR images into 4 tissue classes and assigning predefined attenuation coefficients. For MLAA-GMM, MR images were segmented into known tissue classes, including fat, soft tissue, lung, background air, and an unknown MR low-intensity class encompassing cortical bones, air cavities, and metal artifacts. A coregistered bone probability map was also included in the unknown tissue class. Finally, the GMM prior was constrained over known tissue classes of attenuation maps using unimodal gaussians parameterized over a patient population.
Results: The results showed that the MLAA-GMM algorithm outperformed the MRAC method by differentiating bones from air gaps and providing more accurate patient-specific attenuation coefficients of soft tissue and lungs. It was found that the MRAC and MLAA-GMM methods resulted in average standardized uptake value errors of -5.4% and -3.5% in the lungs, -7.4% and -5.0% in soft tissues/lesions, and -18.4% and -10.2% in bones, respectively.
Conclusion: The proposed MLAA algorithm is promising for accurate derivation of attenuation maps on time-of-flight PET/MR systems.
Keywords: PET/MRI; attenuation correction; quantification; time-of-flight; whole-body imaging.
© 2015 by the Society of Nuclear Medicine and Molecular Imaging, Inc.
Similar articles
-
Joint Estimation of Activity and Attenuation in Whole-Body TOF PET/MRI Using Constrained Gaussian Mixture Models.IEEE Trans Med Imaging. 2015 Sep;34(9):1808-21. doi: 10.1109/TMI.2015.2409157. Epub 2015 Mar 5. IEEE Trans Med Imaging. 2015. PMID: 25769148
-
Quantitative analysis of MRI-guided attenuation correction techniques in time-of-flight brain PET/MRI.Neuroimage. 2016 Apr 15;130:123-133. doi: 10.1016/j.neuroimage.2016.01.060. Epub 2016 Feb 4. Neuroimage. 2016. PMID: 26853602
-
Emission-based estimation of lung attenuation coefficients for attenuation correction in time-of-flight PET/MR.Phys Med Biol. 2015 Jun 21;60(12):4813-33. doi: 10.1088/0031-9155/60/12/4813. Epub 2015 Jun 5. Phys Med Biol. 2015. PMID: 26047036
-
Vision 20/20: Magnetic resonance imaging-guided attenuation correction in PET/MRI: Challenges, solutions, and opportunities.Med Phys. 2016 Mar;43(3):1130-55. doi: 10.1118/1.4941014. Med Phys. 2016. PMID: 26936700 Review.
-
Whole-body FDG PET-MR oncologic imaging: pitfalls in clinical interpretation related to inaccurate MR-based attenuation correction.Abdom Imaging. 2015 Aug;40(6):1374-86. doi: 10.1007/s00261-015-0455-3. Abdom Imaging. 2015. PMID: 26025348 Review.
Cited by
-
PET/MRI: Technical Challenges and Recent Advances.Nucl Med Mol Imaging. 2016 Mar;50(1):3-12. doi: 10.1007/s13139-016-0393-1. Epub 2016 Jan 26. Nucl Med Mol Imaging. 2016. PMID: 26941854 Free PMC article. Review.
-
A review of PET attenuation correction methods for PET-MR.EJNMMI Phys. 2023 Sep 11;10(1):52. doi: 10.1186/s40658-023-00569-0. EJNMMI Phys. 2023. PMID: 37695384 Free PMC article. Review.
-
Direct attenuation correction of brain PET images using only emission data via a deep convolutional encoder-decoder (Deep-DAC).Eur Radiol. 2019 Dec;29(12):6867-6879. doi: 10.1007/s00330-019-06229-1. Epub 2019 Jun 21. Eur Radiol. 2019. PMID: 31227879
-
Time of Flight in Perspective: Instrumental and Computational Aspects of Time Resolution in Positron Emission Tomography.IEEE Trans Radiat Plasma Med Sci. 2021 Sep;5(5):598-618. doi: 10.1109/trpms.2021.3084539. Epub 2021 May 27. IEEE Trans Radiat Plasma Med Sci. 2021. PMID: 34553105 Free PMC article.
-
Deep-JASC: joint attenuation and scatter correction in whole-body 18F-FDG PET using a deep residual network.Eur J Nucl Med Mol Imaging. 2020 Oct;47(11):2533-2548. doi: 10.1007/s00259-020-04852-5. Epub 2020 May 15. Eur J Nucl Med Mol Imaging. 2020. PMID: 32415552
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical