Management of a type Ia endoleak with the Nellix endovascular aneurysm sealing system
- PMID: 25862364
- DOI: 10.1177/1526602815579254
Management of a type Ia endoleak with the Nellix endovascular aneurysm sealing system
Abstract
Purpose: To report the application of the Nellix endovascular aneurysm sealing (EVAS) device, including two chimney grafts, to successfully treat a type Ia endoleak.
Case report: An 87-year-old man had an asymptomatic 7.6-cm infrarenal abdominal aortic aneurysm (AAA) and a 4.5-cm right internal iliac artery aneurysm treated using an aortouni-iliac stent-graft. Two years after the index endovascular repair, an asymptomatic type Ia endoleak was detected on duplex ultrasound; the computed tomographic angiogram (CTA) demonstrated significant sac enlargement and stent-graft migration. Initial attempts to treat the leak with 2 aortic cuffs only reduced the size of the endoleak. Another procedure was undertaken using the Nellix device with chimney grafts to increase the proximal sealing zone above the existing stent-graft. Imaging postoperatively demonstrated successful resolution of the endoleak and continuing patency of both renal artery chimney stent-grafts. CTA at 6 months demonstrated persistent sealing of the endoleak.
Conclusion: The use of the EVAS system may represent another endovascular solution that can be added to the clinician's repertoire for treating type Ia endoleak after conventional endovascular repair of infrarenal AAA.
Keywords: abdominal aortic aneurysm; chimney graft; endobags; endograft; endoleak; endovascular aneurysm repair; endovascular sealing system; reintervention; stent-graft.
© The Author(s) 2015.
Comment in
-
Commentary: endovascular aneurysm sealing to treat proximal type I endoleak: a new method to stanch the leak?J Endovasc Ther. 2015 Jun;22(3):312-3. doi: 10.1177/1526602815582549. Epub 2015 Apr 22. J Endovasc Ther. 2015. PMID: 25904482 No abstract available.
Similar articles
-
An Alternative Approach for Treating a Type Ia Endoleak after Conventional EVAR Using the Nellix Endovascular Aneurysm Sealing.Ann Vasc Surg. 2018 May;49:316.e5-316.e10. doi: 10.1016/j.avsg.2017.12.020. Epub 2018 Mar 1. Ann Vasc Surg. 2018. PMID: 29501907
-
Management of Nellix migration and type Ia endoleak from proximal endovascular aneurysm sealing relining to late open conversion.J Vasc Surg. 2021 Oct;74(4):1204-1213. doi: 10.1016/j.jvs.2021.02.035. Epub 2021 Mar 5. J Vasc Surg. 2021. PMID: 33684472
-
Feasibility and Technical Aspects of Proximal Nellix-in-Nellix Extension for Late Caudal Endograft Migration.J Endovasc Ther. 2017 Apr;24(2):210-217. doi: 10.1177/1526602816677037. Epub 2016 Nov 19. J Endovasc Ther. 2017. PMID: 27864459
-
Percutaneous interventions following endovascular aneurysm sac sealing: Endoleak embolization and limb-related adverse events.Semin Vasc Surg. 2016 Sep;29(3):135-141. doi: 10.1053/j.semvascsurg.2016.11.005. Epub 2016 Nov 5. Semin Vasc Surg. 2016. PMID: 27989319 Review.
-
Endovascular aneurysm sealing as an alternative for the treatment of failed endovascular aneurysm repair.Vascular. 2020 Jun;28(3):251-258. doi: 10.1177/1708538119897525. Epub 2020 Jan 2. Vascular. 2020. PMID: 31896300
Cited by
-
An International, Multicenter Retrospective Observational Study to Assess Technical Success and Clinical Outcomes of Patients Treated with an Endovascular Aneurysm Sealing Device for Type III Endoleak.J Endovasc Ther. 2022 Feb;29(1):57-65. doi: 10.1177/15266028211031933. Epub 2021 Aug 3. J Endovasc Ther. 2022. PMID: 34342235 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources