Effect of Irrigant Characteristics on Lesion Formation After Radiofrequency Energy Delivery Using Ablation Catheters with Actively Cooled Tips
- PMID: 25864402
- DOI: 10.1111/jce.12682
Effect of Irrigant Characteristics on Lesion Formation After Radiofrequency Energy Delivery Using Ablation Catheters with Actively Cooled Tips
Abstract
Background: The delivery of radiofrequency (RF) energy through irrigated ablation catheters may be affected by irrigant osmolarity and by catheter position. We sought to characterize lesion formation characteristics using different irrigants in both open and closed irrigated catheter.
Methods: An ex vivo model consisting of viable bovine myocardium and a submersible load cell was assembled in a circulating saline bath at 37°C. An externally irrigated ablation catheter and a closed irrigated catheter were positioned with 10 g of force in both perpendicular and parallel positions. A series of ablation lesions using different irrigants were delivered using a constant rate of irrigation (30 cc/min) at 50 W. Potential clinical applicability was evaluated in vivo by targeting porcine epicardium with different irrigants during open irrigation ablation and assessing lesion sizes.
Results: Ablation in the perpendicular position produced significantly larger lesions for all irrigants, compared to their respective parallel position ablation. For both open and closed irrigated ablation, half normal saline (HNS) ablation created larger lesions than normal saline (NS), and dextrose water (D5W) lesions were significantly larger than both HNS and NS lesions. Steam pops were mostly observed in the perpendicular position, and the rate of steam pops was statistically higher only for open irrigated D5W, but not for HNS, when compared to NS. Both open and closed irrigated ablation with D5W and HNS in the parallel position created larger lesions than parallel NS ablation without causing more steam pops. In an in vivo porcine model, open irrigated ablation with D5W created larger lesions compared to standard NS irrigation.
Conclusions: In ex vivo and in vivo models, decreased osmolarity and charge density increased RF energy delivery to tissue, resulting in larger lesions for both open and closed irrigated ablations. A perpendicular catheter position created larger lesions across all irrigants for both open and closed irrigation ablation. The incidence of steam pops was observed more frequently with high power open irrigated using D5W, especially if the catheter was in a perpendicular position. Further research is required to evaluate any clinical role for using different irrigants with an externally irrigated catheter.
Keywords: Smart Touch catheter; catheter ablation; contact force; experimental study; irrigated catheter tip; ventricular tachycardia.
© 2015 Wiley Periodicals, Inc.
Similar articles
-
Effects of radiofrequency energy delivered through partially insulated metallic catheter tips on myocardial tissue heating and ablation lesion characteristics.Heart Rhythm. 2015 Mar;12(3):623-630. doi: 10.1016/j.hrthm.2014.11.022. Epub 2014 Nov 20. Heart Rhythm. 2015. PMID: 25460861
-
Catheter Ablation Using Half-Normal Saline and Dextrose Irrigation in an Ovine Ventricular Model.JACC Clin Electrophysiol. 2021 Oct;7(10):1229-1239. doi: 10.1016/j.jacep.2021.05.002. Epub 2021 Jun 30. JACC Clin Electrophysiol. 2021. PMID: 34217664
-
Radiofrequency Ablation Using an Open Irrigated Electrode Cooled With Half-Normal Saline.JACC Clin Electrophysiol. 2017 Oct;3(10):1103-1110. doi: 10.1016/j.jacep.2017.03.006. Epub 2017 May 30. JACC Clin Electrophysiol. 2017. PMID: 29759492
-
Biophysics and clinical utility of irrigated-tip radiofrequency catheter ablation.Expert Rev Med Devices. 2012 Jan;9(1):59-70. doi: 10.1586/erd.11.42. Expert Rev Med Devices. 2012. PMID: 22145841 Review.
-
Advances in irrigated tip catheter technology for treatment of cardiac arrhythmias.Recent Pat Cardiovasc Drug Discov. 2013 Apr;8(1):10-6. doi: 10.2174/1574890111308010003. Recent Pat Cardiovasc Drug Discov. 2013. PMID: 23565718 Review.
Cited by
-
Novel Irrigated Temperature-Controlled Lattice Ablation Catheter for Ventricular Ablation: A Preclinical Multimodality Biophysical Characterization.Circ Arrhythm Electrophysiol. 2019 Nov;12(11):e007661. doi: 10.1161/CIRCEP.119.007661. Epub 2019 Nov 11. Circ Arrhythm Electrophysiol. 2019. PMID: 31707809 Free PMC article.
-
2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.J Interv Card Electrophysiol. 2020 Oct;59(1):145-298. doi: 10.1007/s10840-019-00663-3. J Interv Card Electrophysiol. 2020. PMID: 31984466 Free PMC article.
-
Approaching Ventricular Tachycardia Ablation in 2024: An Update on Mapping and Ablation Strategies, Timing, and Future Directions.J Clin Med. 2024 Aug 24;13(17):5017. doi: 10.3390/jcm13175017. J Clin Med. 2024. PMID: 39274230 Free PMC article. Review.
-
Moving the needle: Tissue characterization and lesion formation during infusion-needle ablation.Heart Rhythm. 2020 Mar;17(3):406-407. doi: 10.1016/j.hrthm.2019.10.018. Epub 2019 Oct 12. Heart Rhythm. 2020. PMID: 31614225 Free PMC article. No abstract available.
-
Nonionic irrigated radiofrequency ablation of refractory incessant ventricular tachycardia via great cardiac vein.HeartRhythm Case Rep. 2018 Sep 5;4(12):572-575. doi: 10.1016/j.hrcr.2018.08.012. eCollection 2018 Dec. HeartRhythm Case Rep. 2018. PMID: 30581735 Free PMC article. No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous