A prospective comparison of performance during back-to-back, anterograde manual spiral enteroscopy and double-balloon enteroscopy

Affiliations
Comparative Study

A prospective comparison of performance during back-to-back, anterograde manual spiral enteroscopy and double-balloon enteroscopy

Edward J Despott et al. Dig Liver Dis. 2015 May.

Abstract

Background: Spiral enteroscopy is a recently introduced technology alternative to balloon-assisted enteroscopy for examination of the small bowel.

Aim: To compare small bowel insertion depths and procedure duration by spiral enteroscopy and double-balloon enteroscopy performed in the same cohort of patients, in immediate succession, using the same method of insertion depth estimation.

Methods: A prospective, back-to-back comparative study was performed in 15 patients. Spiral enteroscopy procedures were performed first and a tattoo was placed to mark the most distal point.

Results: Double-balloon enteroscopy passed the tattoo placed at spiral enteroscopy in 14/15 cases (93%). Median insertion depths for double-balloon enteroscopy and spiral enteroscopy were 265cm and 175cm, respectively (P=0.004). Median time to achieve maximal depth of insertion was significantly shorter for spiral enteroscopy compared with double-balloon enteroscopy (24min vs. 45min, respectively; P=0.0005). However, in 14 patients no differences were found in median time to reach the same insertion depth (P=0.28).

Conclusion: Double-balloon enteroscopy achieved significantly greater small bowel insertion depth than spiral enteroscopy. Although overall double-balloon enteroscopy procedure duration was longer, the time taken to reach the same small bowel insertion depth by both spiral enteroscopy and double-balloon enteroscopy was similar.

Keywords: Depth of insertion; Device assisted enteroscopy; Enteroscopy; Small bowel.