Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial
- PMID: 25881930
 - DOI: 10.1016/j.jacc.2015.02.022
 
Chronic performance of a leadless cardiac pacemaker: 1-year follow-up of the LEADLESS trial
Abstract
Background: A leadless cardiac pacemaker (LCP) system was recently introduced to overcome lead-related complications of conventional pacing systems. To date, long-term results of an LCP system are unknown.
Objectives: The aim of this study was to assess the complication incidence, electrical performance, and rate response characteristics within the first year of follow-up of patients implanted with an LCP.
Methods: We retrospectively assessed intermediate-term follow-up data for 31 of 33 patients from the LEADLESS trial cohort who had an indication for single-chamber pacing and received an LCP between December 2012 and April 2013.
Results: The mean age of the cohort was 76 ± 8 years, and 65% were male. Between 3 and 12 months of follow-up, there were no pacemaker-related adverse events reported. The pacing performance results at 6- and 12-month follow-up were, respectively, as follows: mean pacing threshold (at a 0.4-ms pulse width), 0.40 ± 0.26 V and 0.43 ± 0.30 V; R-wave amplitude 10.6 ± 2.6 mV and 10.3 ± 2.2 mV; and impedance 625 ± 205 Ω and 627 ± 209 Ω. At the 12-month follow-up in 61% of the patients (n = 19 of 31), the rate response sensor was activated, and an adequate rate response was observed in all patients.
Conclusions: The LCP demonstrates very stable performance and reassuring safety results during intermediate-term follow-up. These results support the use of the LCP as a promising alternative to conventional pacemaker systems. Continued evaluation is warranted to further characterize this system. (Evaluation of a New Cardiac Pacemaker; NCT01700244).
Keywords: arrhythmia; leadless pacing; pacing; rhythm disorders.
Copyright © 2015 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Comment in
- 
  
  Are leadless pacemakers a niche or the future of device therapy?J Am Coll Cardiol. 2015 Apr 21;65(15):1505-8. doi: 10.1016/j.jacc.2015.02.021. J Am Coll Cardiol. 2015. PMID: 25881931 No abstract available.
 
Publication types
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous