. 2015 Apr;23(4):199-204.
doi: 10.1007/s12471-015-0669-6.

Managing patients with advisory defibrillator leads: what can we learn from published data?

Affiliations

Managing patients with advisory defibrillator leads: what can we learn from published data?

F A Bracke et al. Neth Heart J. 2015 Apr.

Abstract

Defibrillator lead advisories stir a lot of emotions, both with patients and physicians, and this may influence lead management. We reviewed the literature for a more evidence-based approach to this issue.From the complications of two of the current advisory leads, the Medtronic Sprint Fidelis and St. Jude Riata leads, and the consequences of possible interventions, we can conclude that a restrained approach to premature replacement is appropriate. It may be opportune to replace the leads during a scheduled generator replacement in case of a higher electrical failure rate, in order to prevent future premature interventions.We found no support to extract non-functional advisory leads. In contrast, extraction is often more demanding than anticipated, and the risk substantially exceeds that of simply abandoning the leads.

PubMed Disclaimer

Similar articles

References

    1. Kay GN, Brinker JA, Kawanishi DT. Risks of spontaneous injury and extraction of an active fixation pacemaker lead: report of the Accufix Multicenter Clinical Study and Worldwide Registry. Circulation. 1999;100:2344–52. doi: 10.1161/01.CIR.100.23.2344. - DOI - PubMed
    1. Maisel WH, Hauser RG, Hammill SC. Recommendations from the Heart Rhythm Society task force on lead performance policies and guidelines: developed in collaboration with the American College of Cardiology and the American Heart Association. Heart Rhythm. 2009;6:869–85. doi: 10.1016/j.hrthm.2009.04.024. - DOI - PubMed
    1. Abdelhadi RH, Saba SF, Ellis CR. Independent multicenter study of Riata and Riata ST implantable cardioverter-defibrillator leads. Heart Rhythm. 2013;10:361–5. doi: 10.1016/j.hrthm.2012.10.045. - DOI - PubMed
    1. Hayes D, Freedman R, Curtis AB. Prevalence of externalized conductors in Riata and Riata ST silicone leads: Results from the prospective, multicenter Riata Lead Evaluation Study. Heart Rhythm. 2013;10:1778–82. doi: 10.1016/j.hrthm.2013.08.030. - DOI - PubMed
    1. Parkash R, Exner D, Champagne J. Failure rate of the Riata lead under advisory: a report from the CHRS Device Committee. Heart Rhythm. 2013;10:692–5. doi: 10.1016/j.hrthm.2013.01.018. - DOI - PubMed