A comparative study between Truview(PCD) laryngoscope and Macintosh laryngoscope in viewing glottic opening and ease of intubation: A crossover study
- PMID: 25886338
- PMCID: PMC4258957
- DOI: 10.4103/0259-1162.143152
A comparative study between Truview(PCD) laryngoscope and Macintosh laryngoscope in viewing glottic opening and ease of intubation: A crossover study
Abstract
Background: Truview(PCD) laryngoscope is specially designed to aid positioning of the endotracheal tube as well as to record entry of the tube into glottis. Aim of the study is to compare the view of glottic opening and ease of intubation between Truview(PCD) laryngoscope and Macintosh laryngoscope in patients undergoing general anesthesia (GA).
Methodology: Two hundred patients undergoing elective surgery, under GA were randomized into two groups, Group TV and Group ML. In Group TV, Trueview(PCD) laryngoscope was used initially to visualize the vocal cords for Cormack and Lehane grading (CLG) and to spray the vocal cords with 10% lignocaine. Then the patient was ventilated for 1 min and Macintosh laryngoscope was used to visualize the vocal cords for CLG and proceed with intubation. In Group ML, Macintosh laryngoscope was used initially and later Truview(PCD) laryngoscope. Time to intubation, CLG, number of attempts and hemodynamic parameters were recorded.
Results: Ninety-six and 89 patients had CLG1 visualization when Truview(PCD) laryngoscope was used as 1(st) and 2(nd) device respectively compared to 41 and 68 with Macintosh laryngoscope (P = 0.00). Four patients had CLG 4 visualization with Macintosh laryngoscope that turned out to be grade II visualization with Truview(PCD) laryngoscope (P = 0.00). Mean time taken for intubation with Truview(PCD) and Macintosh laryngoscope was 21.10 ± 5.64 s and 15.79 ± 2.76 s respectively (P = 0.00).
Conclusion: Better visualization with lesser CLG was found with Truview(PCD) laryngoscope but it took longer time for intubation than Macintosh laryngoscope. The hemodynamic response to intubation was significantly less with the use of Truview(PCD) laryngoscope when compared to that of Macintosh laryngoscope.
Keywords: Ease of intubation; TruviewPCD laryngoscope; glottic view.
Conflict of interest statement
References
-
- Henderson JJ, Popat MT, Latto IP, Pearce AC Difficult Airway Society. Difficult Airway Society guidelines for management of the unanticipated difficult intubation. Anaesthesia. 2004;59:675–94. - PubMed
-
- Türkan S, Ates Y, Cuhruk H, Tekdemir I. Should we reevaluate the variables for predicting the difficult airway in anesthesiology? Anesth Analg. 2002;94:1340–4. - PubMed
-
- Aprahamian C, Thompson BM, Finger WA, Darin JC. Experimental cervical spine injury model: Evaluation of airway management and splinting techniques. Ann Emerg Med. 1984;13:584–7. - PubMed
-
- Majernick TG, Bieniek R, Houston JB, Hughes HG. Cervical spine movement during orotracheal intubation. Ann Emerg Med. 1986;15:417–20. - PubMed
-
- Brimacombe J, Keller C, Künzel KH, Gaber O, Boehler M, Pühringer F. Cervical spine motion during airway management: A cinefluoroscopic study of the posteriorly destabilized third cervical vertebrae in human cadavers. Anesth Analg. 2000;91:1274–8. - PubMed
LinkOut - more resources
Full Text Sources
Other Literature Sources