. 2015 Jun;10(6):855-65.
doi: 10.1007/s11548-015-1206-1. Epub 2015 Apr 18.

Towards real-time, tracker-less 3D ultrasound guidance for spine anaesthesia

Affiliations

Towards real-time, tracker-less 3D ultrasound guidance for spine anaesthesia

Mikael Brudfors et al. Int J Comput Assist Radiol Surg. 2015 Jun.

Abstract

Purpose: Epidural needle insertions and facet joint injections play an important role in spine anaesthesia. The main challenge of safe needle insertion is the deep location of the target, resulting in a narrow and small insertion channel close to sensitive anatomy. Recent approaches utilizing ultrasound (US) as a low-cost and widely available guiding modality are promising but have yet to become routinely used in clinical practice due to the difficulty in interpreting US images, their limited view of the internal anatomy of the spine, and/or inclusion of cost-intensive tracking hardware which impacts the clinical workflow.

Methods: We propose a novel guidance system for spine anaesthesia. An efficient implementation allows us to continuously align and overlay a statistical model of the lumbar spine on the live 3D US stream without making use of additional tracking hardware. The system is evaluated in vivo on 12 volunteers.

Results: The in vivo study showed that the anatomical features of the epidural space and the facet joints could be continuously located, at a volume rate of 0.5 Hz, within an accuracy of 3 and 7 mm, respectively.

Conclusions: A novel guidance system for spine anaesthesia has been presented which augments a live 3D US stream with detailed anatomical information of the spine. Results from an in vivo study indicate that the proposed system has potential for assisting the physician in quickly finding the target structure and planning a safe insertion trajectory in the spine.

PubMed Disclaimer

References

    1. Anesthesiol Clin. 2008 Mar;26(1):145-58, vii-viii - PubMed
    1. Anesth Analg. 2002 Aug;95(2):411-6, table of contents - PubMed
    1. Acta Anaesthesiol Scand. 2014 Aug;58(7):875-81 - PubMed
    1. IEEE Trans Biomed Eng. 2014 Oct;61(10):2527-37 - PubMed
    1. Med Image Comput Comput Assist Interv. 2012;15(Pt 3):402-9 - PubMed

Publication types

MeSH terms