Comparison of Aerosol Delivery by Face Mask and Tracheostomy Collar

Affiliations
Comparative Study

Comparison of Aerosol Delivery by Face Mask and Tracheostomy Collar

Alaa A Bugis et al. Respir Care. 2015 Sep.

Abstract

Background: The purpose of this study was to compare the performance of a tracheostomy collar, Wright mask, and aerosol mask attached to a jet nebulizer in facilitating aerosolized medication delivery to the lungs. We also compared albuterol delivery with open versus closed fenestration and determined the effect of inspiratory-expiratory ratio (I:E) on aerosol delivery.

Methods: Albuterol (2.5 mg/3 mL) was administered to an in vitro model consisting of an adult teaching mannequin extrathoracic and upper airway with stoma intubated with an 8-mm fenestrated tracheostomy tube. The cuff was deflated. A collecting filter at the level of the bronchi was connected to a breathing simulator at a tidal volume of 400 mL, breathing frequency of 20 breaths/min, and I:E of 2:1 and 1:2. A jet nebulizer was operated with O2 at 8 L/min. Each interface was tested in triplicate. The flow was discontinued at the end of nebulization. For each test, the nebulizer was attached to a tracheostomy collar with the fenestration open or closed, a Wright mask, or an aerosol mask. Drug was analyzed by spectrophotometry (276 nm). A paired t test and analysis of variance were performed (P < .05).

Results: The mean ± SD percent albuterol dose delivered distal to the bronchi was greater with the tracheostomy collar with a closed fenestration (9.4 ± 1.5%) compared with an open fenestration (7.0 ± 0.8%). The doses delivered with the Wright mask (4.1 ± 0.6%) and aerosol mask (3.5 ± 0.04%) were both less than with the tracheostomy collar under either condition (P < .05). Increasing the I:E from 1:2 to 2:1 increased aerosol delivery by 2.5-4%, with significance for the tracheostomy collar with an open fenestration (11.6 ± 1.4%), Wright mask (7.2 ± 0.6%), and aerosol mask (6.1 ± 0.5%).

Conclusions: In an adult tracheostomy model, the tracheostomy collar delivered more aerosol to the bronchi than the Wright or aerosol mask. An I:E of 2:1 caused greater aerosol deposition compared with an I:E of 1:2. During aerosol administration via a tracheostomy collar, closing the fenestration improved aerosol delivery.

Keywords: I:E ratio; aerosols; fenestration; mask; nebulizers; respiratory drug administration; tracheostomy.