Interpolated average CT for cardiac PET/CT attenuation correction
- PMID: 25933679
- DOI: 10.1007/s12350-015-0140-5
Interpolated average CT for cardiac PET/CT attenuation correction
Abstract
Background: Previously, we proposed interpolated averaged CT (IACT) for improved attenuation correction (AC) in thoracic PET/CT. This study aims to evaluate its feasibility and effectiveness on cardiac PET/CT.
Methods: We simulated (18)F-FDG distribution using the XCAT phantom with normal and abnormal cardiac uptake. Average activity and attenuation maps represented static PET and respiration average CT (ACT), respectively, while the attenuation maps of end-inspiration/expiration represented 2 helical CTs (HCT). IACT was obtained by averaging the 2 extreme phases and the interpolated phases generated between them. Later, we recruited 4 patients who were scanned 1 hr post 315-428 MBq (18)F-FDG injection. Simulated and clinical PET sinograms were reconstructed with AC using (1) HCT, (2) IACT, and (3) ACT. Polar plots and the 17-segment plots were analyzed. Two regions-of-interest were drawn on lesion and background area to obtain the intensity ratio (IR).
Results: Polar plots of PETIACT-AC were more similar to PETACT-AC in both simulation and clinical data. Artifacts were observed in various segments in PETHCT-AC. IR differences of HCT as compared to the phantom were up to ~20%.
Conclusions: IACT-AC reduced respiratory artifacts and improved PET/CT matching similarly to ACT-AC. It is a promising low-dose alternate of ACT for cardiac PET/CT.
Keywords: PET/CT; attenuation correction; cardiac imaging; respiratory artifacts.
Comment in
-
Minimizing the radiation dose of CT attenuation correction while improving image quality: The case for innovation.J Nucl Cardiol. 2016 Oct;23(5):1080-1085. doi: 10.1007/s12350-015-0182-8. Epub 2015 Jun 23. J Nucl Cardiol. 2016. PMID: 26100578 No abstract available.
Similar articles
-
Interpolated average CT for attenuation correction in PET--a simulation study.IEEE Trans Biomed Eng. 2013 Jul;60(7):1927-34. doi: 10.1109/TBME.2013.2245132. Epub 2013 Feb 6. IEEE Trans Biomed Eng. 2013. PMID: 23392338
-
Respiratory average CT for attenuation correction in myocardial perfusion SPECT/CT.Ann Nucl Med. 2017 Feb;31(2):172-180. doi: 10.1007/s12149-016-1144-1. Epub 2016 Dec 20. Ann Nucl Med. 2017. PMID: 28000164
-
Interpolated average CT for PET attenuation correction in different lesion characteristics.Nucl Med Commun. 2016 Mar;37(3):297-306. doi: 10.1097/MNM.0000000000000435. Nucl Med Commun. 2016. PMID: 26550984
-
X-ray-based attenuation correction for positron emission tomography/computed tomography scanners.Semin Nucl Med. 2003 Jul;33(3):166-79. doi: 10.1053/snuc.2003.127307. Semin Nucl Med. 2003. PMID: 12931319 Review.
-
Clinical decision support systems in myocardial perfusion imaging.J Nucl Cardiol. 2014 Jun;21(3):427-39; quiz 440. doi: 10.1007/s12350-014-9857-9. Epub 2014 Jan 31. J Nucl Cardiol. 2014. PMID: 24482142 Review.
Cited by
-
Simultaneous reconstruction of attenuation and activity in cardiac PET can remove CT misalignment artifacts.J Nucl Cardiol. 2016 Oct;23(5):1086-1097. doi: 10.1007/s12350-015-0239-8. Epub 2015 Aug 15. J Nucl Cardiol. 2016. PMID: 26275447
-
The effects of mismatch between SPECT and CT images on quantitative activity estimation - A simulation study.Z Med Phys. 2023 Feb;33(1):54-69. doi: 10.1016/j.zemedi.2022.03.004. Epub 2022 May 27. Z Med Phys. 2023. PMID: 35644776 Free PMC article.
-
Pitfalls on PET/CT Due to Artifacts and Instrumentation.Semin Nucl Med. 2021 Nov;51(6):646-656. doi: 10.1053/j.semnuclmed.2021.06.015. Epub 2021 Jul 7. Semin Nucl Med. 2021. PMID: 34243906 Free PMC article. Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials