Randomized Controlled Trial
. 2015 Dec;38(6):1458-67.
doi: 10.1007/s00270-015-1108-1. Epub 2015 May 12.

Value of C-Arm Computed Tomography to Evaluate Stent Deployment During Femoro-Popliteal Revascularization

Affiliations
Randomized Controlled Trial

Value of C-Arm Computed Tomography to Evaluate Stent Deployment During Femoro-Popliteal Revascularization

Gerald Gahide et al. Cardiovasc Intervent Radiol. 2015 Dec.

Abstract

Purpose: To compare the accuracy of C-arm computed tomography (CT) and digital subtraction angiography (DSA) in detecting incomplete stent expansion (ISE) after superficial femoral artery (SFA) stenting using intravascular ultrasound (IVUS) as a gold standard.

Materials: Fifty patients with symptomatic SFA occlusive disease requiring angioplasty were prospectively included. Once technical success (<30 % residual stenosis) was obtained on post-procedural DSA, C-arm CT and IVUS were acquired. DSA and C-arm CT examinations were reviewed by 2 investigators and correlated with IVUS. C-arm CT image quality was rated on a four-point scale. Doppler ultrasound was performed at 1-year follow-up.

Results: The ankle-brachial index was 0.69 ± 0.10 and 0.99 ± 0.40, respectively, pre- and post-procedure. C-arm CT imaging quality was rated as good or excellent in 80%. In-stent minimal luminal diameter (MLD) was evaluated at 4.71 ± 0.7 mm on DSA, 3.39 ± 0.6 mm on IVUS, and 3.12 ± 0.9 mm on C-arm CT. Compared to IVUS, DSA demonstrated an overestimation of MLD (p = 0.0001), an underestimation of ISE (DSA = 18.8% ± 7.6; IVUS = 29.8% ± 9) (p < 0.0001), and a poor inter-technique intra-class correlation coefficient (ICC = 0.24). No difference was observed between IVUS and C-arm CT in ISE as calculated by diameter (29.8 ± 9 vs. 28.2 ± 12.5%, p = 0.5) and area (30.2 ± 8.4 vs. 33.3 ± 9.5%, p = 0.2). Inter-technique ICC between C-arm CT and IVUS was 0.72 [95%CI 0.49; 0.85] for MLA measurements. The inter-observer ICC for MLD and MLA measurements on C-arm CT were, respectively, estimated at 0.75 [95% CI 0.40, 0.89] and 0.77 [95% CI 0.43, 0.90)].

Conclusions: C-arm CT presents a better correlation with IVUS than DSA to determine lumen diameter and ISE immediately after percutaneous revascularization.

Keywords: Imaging; Peripheral vascular disease; Revascularization.