Assessment of four protocols for rapid bacterial identification from positive blood culture pellets by matrix-assisted laser desorption ionization-time of flight mass spectrometry (Vitek® MS)
- PMID: 25966643
- DOI: 10.1016/j.mimet.2015.05.010
Assessment of four protocols for rapid bacterial identification from positive blood culture pellets by matrix-assisted laser desorption ionization-time of flight mass spectrometry (Vitek® MS)
Abstract
In this study, we developed and compared four protocols to prepare a bacterial pellet from 944 positive blood cultures for direct MALDI-TOF mass spectrometry Vitek® MS analysis. Protocol 4, tested on 200 monomicrobial samples, allowed 83% of bacterial identification. This easy, fast, cheap and accurate method is promising in daily practice, especially to limit broad range antibiotic treatment.
Keywords: Blood bottle culture; MALDI-TOF; Rapid identification.
Copyright © 2015 Elsevier B.V. All rights reserved.