Entry zone of iliac screw fixation to maintain proper entry width and screw length
- PMID: 26002356
- DOI: 10.1007/s00586-015-4042-7
Entry zone of iliac screw fixation to maintain proper entry width and screw length
Abstract
Purpose: To evaluate the entry zone of iliac screw fixation to maintain proper entry width and screw length.
Methods: Computed tomography images of pelvic bones from 90 human cadavers were reconstructed into 3-dimensional models. In each model, a sectional image crossing the posterior superior iliac spine (PSIS) and anterior inferior iliac spine (AIIS) and consecutive sectional images up to 20 mm superiorly and inferiorly from the PSIS with 1-mm intervals aiming the AIIS were obtained. One virtual iliac screw with 10-mm diameter was introduced onto the PSIS at the middle and at the lateral and medial 1/4 points on the prominence of the posterior iliac spine. The entry width of the bony prominence and the corresponding maximal screw length available were evaluated for each entry point.
Results: The entry width was smallest on the inferior 20 mm (4.7 ± 3.0 mm) and gradually increased up to the superior 10 mm (19.1 ± 3.9 mm) sectional images. The maximal screw length was smallest on the superior 20 mm (76.7 ± 39.7 mm) and gradually increased down to the inferior 10 mm (112.3 ± 15.1 mm) sectional images. The maximal screw lengths were significantly greatest at the most medial point and smallest at the most lateral point on the superior 20- and 10-mm sectional images and at the PSIS.
Conclusions: The iliac screw fixation entry zone to maintain proper screw length and entry width is outlined from 20 mm superiorly to 10 mm inferiorly from the PSIS and is located more medially from the prominence of the posterior iliac spine.
Keywords: Entry point; Entry zone; Iliac screw; Maximal screw length; Spinopelvic fixation.
Similar articles
-
Low profile pelvic fixation: anatomic parameters for sacral alar-iliac fixation versus traditional iliac fixation.Spine (Phila Pa 1976). 2009 Mar 1;34(5):436-40. doi: 10.1097/BRS.0b013e318194128c. Spine (Phila Pa 1976). 2009. PMID: 19247163
-
Effect of iliac screw insertion depth on the stability and strength of lumbo-iliac fixation constructs: an anatomical and biomechanical study.Spine (Phila Pa 1976). 2009 Jul 15;34(16):E565-72. doi: 10.1097/BRS.0b013e3181ac8fc4. Spine (Phila Pa 1976). 2009. PMID: 19770599
-
Safe insertion of S-2 alar iliac screws: radiological comparison between 2 insertion points using computed tomography and 3D analysis software.J Neurosurg Spine. 2018 May;28(5):536-542. doi: 10.3171/2017.8.SPINE17735. Epub 2018 Feb 16. J Neurosurg Spine. 2018. PMID: 29451435
-
Optimizing iliac screw fixation: a biomechanical study on screw length, trajectory, and diameter.J Neurosurg Spine. 2011 Feb;14(2):219-25. doi: 10.3171/2010.9.SPINE10254. Epub 2010 Dec 24. J Neurosurg Spine. 2011. PMID: 21184640
-
Morphometry of iliac anchorage for transiliac screws: a cadaver and CT study of the Eastern population.Surg Radiol Anat. 2010 Jun;32(5):455-62. doi: 10.1007/s00276-009-0589-5. Epub 2009 Nov 20. Surg Radiol Anat. 2010. PMID: 19936596
Cited by
-
Comparative tomographic study of the S2-alar-iliac screw versus the iliac screw.Eur Spine J. 2019 Apr;28(4):855-862. doi: 10.1007/s00586-018-5806-7. Epub 2018 Oct 31. Eur Spine J. 2019. PMID: 30382431
-
Comparative tomographic study of the iliac screw and the S2-alar-iliac screw in children.Clinics (Sao Paulo). 2020 Sep 11;75:e1824. doi: 10.6061/clinics/2020/e1824. eCollection 2020. Clinics (Sao Paulo). 2020. PMID: 32935824 Free PMC article.
-
Technical variation of trans-articular sacroiliac joint (SIJ) fusion using three screws considering the effects of sacral dysplasia in patients with non-traumatic SIJ pain.BMC Musculoskelet Disord. 2019 Aug 28;20(1):386. doi: 10.1186/s12891-019-2771-1. BMC Musculoskelet Disord. 2019. PMID: 31455346 Free PMC article.
-
Anatomic Parameters for Instrumentation of the Sacrum and Pelvis: A Systematic Review of the Literature.J Am Acad Orthop Surg Glob Res Rev. 2018 Aug 2;2(8):e034. doi: 10.5435/JAAOSGlobal-D-18-00034. eCollection 2018 Aug. J Am Acad Orthop Surg Glob Res Rev. 2018. PMID: 30631829 Free PMC article.
-
CT-based Anatomic and Clinical Analysis of Iliac Screw Placement During Spinopelvic Fixation.In Vivo. 2025 May-Jun;39(3):1514-1523. doi: 10.21873/invivo.13951. In Vivo. 2025. PMID: 40295015 Free PMC article.