. 2015 May-Jun;12(3):271-7.

Accuracy and reliability of linear measurements using tangential projection and cone beam computed tomography

Affiliations

Accuracy and reliability of linear measurements using tangential projection and cone beam computed tomography

Mahnaz Sheikhi et al. Dent Res J (Isfahan). 2015 May-Jun.

Abstract

Background: Providing a cross-sectional image is essential for preimplant assessments. Computed tomography (CT) and cone beam CT (CBCT) images are very expensive and provide high radiation dose. Tangential projection is a very simple, available, and low-dose technique that can be used in the anterior portion of mandible. The purpose of this study was to evaluate the accuracy of tangential projection in preimplant measurements in comparison to CBCT.

Materials and methods: Three dry edentulous human mandibles were examined in five points at intercanine region using tangential projection and CBCT. The height and width of the ridge were measured twice by two observers. The mandibles were then cut, and real measurements were obtained. The agreement between real measures and measurements obtained by either technique, and inter- and intra-observer reliability were tested.

Results: The measurement error was less than 0.12 for tangential technique and 0.06 for CBCT. The agreement between the real measures and measurements from radiographs were higher than 0.87. Tangential projection slightly overestimated the distances, while there was a slight underestimation in CBCT results.

Conclusion: Considering the low cost, low radiation dose, simplicity and availability, tangenital projection would be adequate for preimplant assessment in edentulous patients when limited numbers of implants are required in the anterior mandible.

Keywords: Cone beam computed tomography; dental implants; radiography.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: The authors of this manuscript declare that they have no conflicts of interest, real or perceived, financial or non-financial in this article.

Figures

Figure 1
Figure 1
The fabricated film holder, the container, and a specimen positioned in it.
Figure 2
Figure 2
(a) cross sectional CBCT image, (b) tangential projection provide by the film holder.
Figure 3
Figure 3
The interclass correlation co-efficience between the measurements of the length in tangential projection and bone.
Figure 4
Figure 4
The interclass correlation co-efficience between measurements of the length in cone beam computed tomography and bone.
Figure 5
Figure 5
The interclass correlation co-efficience between measurements of the width in tangential projection and bone.
Figure 6
Figure 6
The interclass correlation co-efficience between measurements of the width in cone beam computed tomography and bone.

Similar articles

Cited by

References

    1. Loubele M, Van Assche N, Carpentier K, Maes F, Jacobs R, van Steenberghe D, et al. Comparative localized linear accuracy of small-field cone-beam CT and multislice CT for alveolar bone measurements. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 2008;105:512–8. - PubMed
    1. Gupta J, Ali SP. Cone beam computed tomography in oral implants. Natl J Maxillofac Surg. 2013;4:2–6. - PMC - PubMed
    1. Sheikhi M, Ghorbanizadeh S, Abdinian M, Goroohi H, Badrian H. Accuracy of linear measurements of galileos cone beam computed tomography in normal and different head positions. Int J Dent 2012. 2012 214954. - PMC - PubMed
    1. Monsour PA, Dudhia R. Implant radiography and radiology. Aust Dent J. 2008;53(Suppl 1):S11–25. - PubMed
    1. Hauret L, Hodez C. A new modality for dentomaxillofacial imaging: Cone beam CT. J Radiol. 2009;90:604–17. - PubMed