Accuracy and reliability of linear measurements using tangential projection and cone beam computed tomography
- PMID: 26005469
- PMCID: PMC4432612
Accuracy and reliability of linear measurements using tangential projection and cone beam computed tomography
Abstract
Background: Providing a cross-sectional image is essential for preimplant assessments. Computed tomography (CT) and cone beam CT (CBCT) images are very expensive and provide high radiation dose. Tangential projection is a very simple, available, and low-dose technique that can be used in the anterior portion of mandible. The purpose of this study was to evaluate the accuracy of tangential projection in preimplant measurements in comparison to CBCT.
Materials and methods: Three dry edentulous human mandibles were examined in five points at intercanine region using tangential projection and CBCT. The height and width of the ridge were measured twice by two observers. The mandibles were then cut, and real measurements were obtained. The agreement between real measures and measurements obtained by either technique, and inter- and intra-observer reliability were tested.
Results: The measurement error was less than 0.12 for tangential technique and 0.06 for CBCT. The agreement between the real measures and measurements from radiographs were higher than 0.87. Tangential projection slightly overestimated the distances, while there was a slight underestimation in CBCT results.
Conclusion: Considering the low cost, low radiation dose, simplicity and availability, tangenital projection would be adequate for preimplant assessment in edentulous patients when limited numbers of implants are required in the anterior mandible.
Keywords: Cone beam computed tomography; dental implants; radiography.
Conflict of interest statement
Figures






Similar articles
-
Reliability and accuracy of cone beam computed tomography versus conventional multidetector computed tomography for image-guided craniofacial implant planning: An in vitro study.Int J Oral Maxillofac Implants. 2019 May/June;34(3):665–672. doi: 10.11607/jomi.6915. Epub 2019 Apr 1. Int J Oral Maxillofac Implants. 2019. PMID: 30934042
-
Accuracy of linear measurements using cone beam computed tomography and panoramic radiography in dental implant treatment planning.Int J Oral Maxillofac Implants. 2015 Nov-Dec;30(6):1287-94. doi: 10.11607/jomi.4073. Int J Oral Maxillofac Implants. 2015. PMID: 26574854
-
Accuracy of linear measurements using dental cone beam and conventional multislice computed tomography.Dentomaxillofac Radiol. 2008 Jan;37(1):10-7. doi: 10.1259/dmfr/14140281. Dentomaxillofac Radiol. 2008. PMID: 18195249
-
Accuracy of linear measurements on CBCT images related to presurgical implant treatment planning: A systematic review.Clin Oral Implants Res. 2018 Oct;29 Suppl 16:393-415. doi: 10.1111/clr.13142. Clin Oral Implants Res. 2018. PMID: 30328204
-
A comparison of jaw dimensional and quality assessments of bone characteristics with cone-beam CT, spiral tomography, and multi-slice spiral CT.Int J Oral Maxillofac Implants. 2007 May-Jun;22(3):446-54. Int J Oral Maxillofac Implants. 2007. PMID: 17622012
Cited by
-
Comparison of digital panoramic radiography versus cone beam computerized tomography for measuring alveolar bone.Head Face Med. 2017 Feb 22;13(1):2. doi: 10.1186/s13005-017-0135-3. Head Face Med. 2017. PMID: 28228135 Free PMC article.
-
Influence of voxel size on the accuracy of linear measurements of the condyle in images of cone beam computed tomography: A pilot study.J Clin Exp Dent. 2018 Sep 1;10(9):e876-e882. doi: 10.4317/jced.54500. eCollection 2018 Sep. J Clin Exp Dent. 2018. PMID: 30386520 Free PMC article.
-
Comparing the precision of panoramic radiography and cone-beam computed tomography in avoiding anatomical structures critical to dental implant surgery: A retrospective study.Imaging Sci Dent. 2018 Dec;48(4):269-275. doi: 10.5624/isd.2018.48.4.269. Epub 2018 Dec 20. Imaging Sci Dent. 2018. PMID: 30607351 Free PMC article.
-
Intraobserver and interobserver reproducibility in linear measurements on axial images obtained by cone-beam computed tomography.Imaging Sci Dent. 2017 Mar;47(1):11-15. doi: 10.5624/isd.2017.47.1.11. Epub 2017 Mar 21. Imaging Sci Dent. 2017. PMID: 28361024 Free PMC article.
-
Comparison of Cone Beam Computed Tomography and Multi Slice Computed Tomography Image Quality of Human Dried Mandible using 10 Anatomical Landmarks.J Clin Diagn Res. 2017 Feb;11(2):ZC13-ZC16. doi: 10.7860/JCDR/2017/20637.9253. Epub 2017 Feb 1. J Clin Diagn Res. 2017. PMID: 28384972 Free PMC article.