In vitro fatigue and fracture resistance of one- and two-piece CAD/CAM zirconia implant abutments
- PMID: 26009905
- DOI: 10.11607/jomi.3942
In vitro fatigue and fracture resistance of one- and two-piece CAD/CAM zirconia implant abutments
Abstract
Purpose: All-ceramic abutments are employed increasingly often in implant dentistry for esthetic reasons. In vitro stress testing is required to evaluate the suitability of these constructions, especially in load-bearing posterior regions. The purpose of the study was to assess and compare the fatigue and fracture resistance of one- and two-piece computer-aided design/computer-assisted manufacture (CAD/CAM) zirconia implant abutments with an internal-hex connection and prefabricated commercially available zirconia stock abutments.
Materials and methods: Twenty-one abutment-crown specimens were prepared for three test groups. Control group 1 (SZ) included specimens with unprepared stock zirconia abutments, test group 2 (OP) included one-piece CAD/CAM zirconia abutments, and test group 3 (TP) included two-piece CAD/CAM zirconia abutments. All 21 specimens underwent thermocycling and fatigue testing. Finally, all specimens were tested for fracture resistance with a universal testing machine. The maximum load was applied to the tapered occlusal area of each crown at a 30-degree angle and a crosshead speed of 0.5 mm/min until the implant-abutment connection failed. Kolmogorov-Smirnov, Shapiro-Wilk, and post-hoc Scheffé tests were used for statistical analysis.
Results: All abutments in groups SZ and OP fractured into two or more pieces after fracture resistance testing. None of the TP abutments displayed apparent disintegration, but failure was evidenced by bending of the retention screw. OP abutments (232.1 ± 29.8 N) and SZ abutments (251.8 ± 23.2 N) showed lower fracture loads than the TP abutments (291.4 ± 27.8 N). However, only the difference between the OP and TP groups was statistically significant. Further load-displacement analyses corroborated the higher mechanical stability of the TP abutments.
Conclusion: Superior resistance was achieved for two-piece hybrid CAD/CAM zirconia abutments. These abutments might be clinically beneficial in high-load areas, such as premolar and molar regions.
Similar articles
-
Bending moments of zirconia and titanium implant abutments supporting all-ceramic crowns after aging.Clin Oral Implants Res. 2014 Jan;25(1):74-81. doi: 10.1111/clr.12192. Epub 2013 Jun 4. Clin Oral Implants Res. 2014. PMID: 23735182
-
Mechanical stability of fully personalized, abutment-free zirconia implant crowns on a novel implant-crown interface.J Dent. 2022 Jun;121:104121. doi: 10.1016/j.jdent.2022.104121. Epub 2022 Apr 5. J Dent. 2022. PMID: 35395344
-
Retentive strength of two-piece CAD/CAM zirconia implant abutments.Clin Implant Dent Relat Res. 2014 Dec;16(6):920-5. doi: 10.1111/cid.12060. Epub 2013 Mar 25. Clin Implant Dent Relat Res. 2014. PMID: 23527950
-
Zirconia in fixed implant prosthodontics.Clin Implant Dent Relat Res. 2012 Oct;14(5):633-45. doi: 10.1111/j.1708-8208.2010.00317.x. Epub 2010 Dec 22. Clin Implant Dent Relat Res. 2012. PMID: 21176095 Review.
-
Fracture of Zirconia Abutments in Implant Treatments: A Systematic Review.Implant Dent. 2019 Aug;28(4):378-387. doi: 10.1097/ID.0000000000000900. Implant Dent. 2019. PMID: 31344017
Cited by
-
The biocompatibility and mechanical properties of plasma sprayed zirconia coated abutment.J Adv Prosthodont. 2020 Jun;12(3):157-166. doi: 10.4047/jap.2020.12.3.157. Epub 2020 Jun 18. J Adv Prosthodont. 2020. PMID: 32601535 Free PMC article.
-
Effect of cement type, luting protocol, and ceramic abutment material on the shade of cemented titanium-based lithium disilicate crowns and surrounding peri-implant soft tissue: a spectrophotometric analysis.J Adv Prosthodont. 2024 Aug;16(4):231-243. doi: 10.4047/jap.2024.16.4.231. Epub 2024 Aug 20. J Adv Prosthodont. 2024. PMID: 39221417 Free PMC article.
-
Effects of Liner-Bonding of Implant-Supported Glass-Ceramic Crown to Zirconia Abutment on Bond Strength and Fracture Resistance.Materials (Basel). 2019 Aug 30;12(17):2798. doi: 10.3390/ma12172798. Materials (Basel). 2019. PMID: 31480271 Free PMC article.
-
A Comparative Study on the Mechanical Properties of a Polymer-Infiltrated Ceramic-Network Material Used for the Fabrication of Hybrid Abutment.Materials (Basel). 2018 Sep 11;11(9):1681. doi: 10.3390/ma11091681. Materials (Basel). 2018. PMID: 30208618 Free PMC article.
-
New Ceramic Multi-Unit Dental Abutments with an Antimicrobial Glassy Coating.Materials (Basel). 2022 Aug 6;15(15):5422. doi: 10.3390/ma15155422. Materials (Basel). 2022. PMID: 35955356 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous