. 2015 Jun;29(3):178-84.
doi: 10.3341/kjo.2015.29.3.178. Epub 2015 May 20.

Comparison of the refractive measurements with hand-held autorefractometer, table-mounted autorefractometer and cycloplegic retinoscopy in children

Affiliations

Comparison of the refractive measurements with hand-held autorefractometer, table-mounted autorefractometer and cycloplegic retinoscopy in children

Handan Akil et al. Korean J Ophthalmol. 2015 Jun.

Abstract

Purpose: To evaluate the performance of the hand-held and table-top autorefractokeratometer in measuring refractive errors by comparing them with cycloplegic retinoscopy.

Methods: Included in the study were 112 eyes of 112 pediatric patients whose mean age was 6.78 ± 2.61 years (range, 2 to 12 years). The refractive errors of all the eyes were measured with and without cycloplegia using a hand held autorefractokeratometer (Retinomax K-plus 3), table top autorefractokeratometer (Canon RK-F1) and performing cycloplegic retinoscopy. The spherical equivalent, cylindrical axis and keratometer values were statistically compared.

Results: The mean spherical equivalent obtained from the Retinomax K-plus 3 was significantly less hyperopic than that of Canon RK-F1 (p = 0.004) before cycloplegia. When the Bland Altman analysis was performed in comparisons of spherical equivalent values measured with the Retinomax K-plus 3, Canon RK-F1 and cycloplegic retinoscopy, it was seen that almost all of the differences between the measurements remained within the range of ±2 standard deviation. Good agreement was found between Retinomax K-plus 3 and Canon RK-F1 for the Jackson cross-cylinder values at axis 0° and 45°; keratometer values respectively.

Conclusions: The refractive error components were highly correlated between the two instruments and cycloplegic retinoscopy.

Keywords: Hand-held autorefractokeratometer; Refractive errors; Retinoscopy; Table-top autorefractokeratometer.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: No potential conflict of interest relevant to this article was reported.

References

    1. Tongue AC. Refractive errors in children. Pediatr Clin North Am. 1987;34:1425–1437. - PubMed
    1. Atkinson J, Braddick O, Robier B, et al. Two infant vision screening programmes: prediction and prevention of strabismus and amblyopia from photo- and videorefractive screening. Eye (Lond) 1996;10(Pt 2):189–198. - PubMed
    1. Lithander J, Sjostrand J. Anisometropic and strabismic amblyopia in the age group 2 years and above: a prospective study of the results of treatment. Br J Ophthalmol. 1991;75:111–116. - PMC - PubMed
    1. Schmidt P, Maguire M, Dobson V, et al. Comparison of preschool vision screening tests as administered by licensed eye care professionals in the Vision In Preschoolers Study. Ophthalmology. 2004;111:637–650. - PubMed
    1. Williams C, Lumb R, Harvey I, Sparrow JM. Screening for refractive errors with the Topcon PR2000 Pediatric Refractometer. Invest Ophthalmol Vis Sci. 2000;41:1031–1037. - PubMed

LinkOut - more resources