Measuring motion-induced B0 -fluctuations in the brain using field probes
Abstract
Purpose: Fluctuations of the background magnetic field (B0 ) due to body and breathing motion can lead to significant artifacts in brain imaging at ultrahigh field. Corrections based on real-time sensing using external field probes show great potential. This study evaluates different aspects of field interpolation from these probes into the brain which is implicit in such methods. Measurements and simulations were performed to quantify how well B0 -fluctuations in the brain due to body and breathing motion are reflected in external field probe measurements.
Methods: Field probe measurements were compared with scanner acquired B0 -maps from experiments with breathing and shoulder movements. A realistic simulation of B0 -fluctuations caused by breathing was performed, and used for testing different sets of field probe positions.
Results: The B0 -fluctuations were well reflected in the field probe measurements in the shoulder experiments, while the breathing experiments showed only moderate correspondence. The simulations showed the importance of the probe positions, and that performing full 3(rd) order corrections based on 16 field probes is not recommended.
Conclusion: Methods for quantitative assessment of the field interpolation problem were developed and demonstrated. Field corrections based on external field measurements show great potential, although potential pitfalls were identified.
Keywords: 7 Tesla; NMR field probes; dynamic shimming; field fluctuations; field mapping; motion.
© 2015 Wiley Periodicals, Inc.
Similar articles
-
Real-time shimming to compensate for respiration-induced B0 fluctuations.Magn Reson Med. 2007 Feb;57(2):362-8. doi: 10.1002/mrm.21136. Magn Reson Med. 2007. PMID: 17260378
-
Direct B0 field monitoring and real-time B0 field updating in the human breast at 7 Tesla.Magn Reson Med. 2012 Feb;67(2):586-91. doi: 10.1002/mrm.23272. Epub 2011 Dec 12. Magn Reson Med. 2012. PMID: 22161736
-
Rapid measurement and correction of spatiotemporal B0 field changes using FID navigators and a multi-channel reference image.Magn Reson Med. 2020 Feb;83(2):575-589. doi: 10.1002/mrm.27957. Epub 2019 Aug 29. Magn Reson Med. 2020. PMID: 31463976 Free PMC article.
-
A comparison of navigators, snap-shot field monitoring, and probe-based field model training for correcting B0 -induced artifacts in T2*-weighted images at 7 T.Magn Reson Med. 2017 Oct;78(4):1373-1382. doi: 10.1002/mrm.26524. Epub 2016 Nov 17. Magn Reson Med. 2017. PMID: 27859614
-
B0 magnetic field homogeneity and shimming for in vivo magnetic resonance spectroscopy.Anal Biochem. 2017 Jul 15;529:17-29. doi: 10.1016/j.ab.2016.06.003. Epub 2016 Jun 9. Anal Biochem. 2017. PMID: 27293215 Free PMC article. Review.
Cited by
-
Dynamic distortion correction for functional MRI using FID navigators.Magn Reson Med. 2021 Mar;85(3):1294-1307. doi: 10.1002/mrm.28505. Epub 2020 Sep 24. Magn Reson Med. 2021. PMID: 32970869 Free PMC article.
-
Inductive measurement and encoding of k-space trajectories in MR raw data.MAGMA. 2019 Dec;32(6):655-667. doi: 10.1007/s10334-019-00770-2. Epub 2019 Jul 30. MAGMA. 2019. PMID: 31363869
-
Motion correction methods for MRS: experts' consensus recommendations.NMR Biomed. 2021 May;34(5):e4364. doi: 10.1002/nbm.4364. Epub 2020 Jul 20. NMR Biomed. 2021. PMID: 33089547 Free PMC article.
-
Real-time shimming with FID navigators.Magn Reson Med. 2022 Dec;88(6):2548-2563. doi: 10.1002/mrm.29421. Epub 2022 Sep 12. Magn Reson Med. 2022. PMID: 36093989 Free PMC article.
-
Feasibility of 31 P spectroscopic imaging at 7 T in lung carcinoma patients.NMR Biomed. 2021 May;34(5):e4204. doi: 10.1002/nbm.4204. Epub 2019 Nov 17. NMR Biomed. 2021. PMID: 31736167 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous