. 2015 Sep;62(3):578-84.
doi: 10.1016/j.jvs.2015.04.395. Epub 2015 Jun 16.

Use of parallel grafts to save failed prior endovascular aortic aneurysm repair and type Ia endoleaks

Affiliations
Free article

Use of parallel grafts to save failed prior endovascular aortic aneurysm repair and type Ia endoleaks

Konstantinos P Donas et al. J Vasc Surg. 2015 Sep.
Free article

Abstract

Objective: The aim of this study was the evaluation of the clinical and radiologic outcomes of parallel grafts in the treatment of patients with failed prior endovascular aneurysm repair and type Ia endoleak.

Methods: Prospectively collected clinical and radiologic data of consecutive patients with prior endovascular aneurysm repair and evidence of type Ia endoleak were analyzed. All patients were treated between January 2009 and November 2014 by use of parallel grafts, ie, chimney/snorkel or periscope grafts and abdominal endovascular devices. Primary outcome of the study was the technical success. Secondary outcomes were patency of the chimney grafts and shrinkage of aneurysm morphology.

Results: Technical success was 94.4% (17 of 18). One patient showed evidence of persistent type Ia endoleak after triple chimney graft placement and treated by Onyx embolization (Covidien, Plymouth, Minn) successfully. The mean preoperative and postoperative aneurysm diameters were 7.1 ± 1.7 and 7.0 ± 2 cm. Primary patency of the chimney grafts was 96.7% and assisted patency was 100% after successful retrograde recanalization of an acute occlusion of the superior mesenteric artery from the celiac trunk and the gastroduodenal artery. One patient who had undergone the sandwich technique with transfemoral placement of the flexible Viabahn (W. L. Gore & Associates, Flagstaff, Ariz) due to impossible catheterization of the target vessel from the upper extremity presented with contained rupture caused by gutter-associated endoleak and underwent urgent embolization with Onyx successfully. No procedure-related death or open conversion was observed in the entire follow-up period.

Conclusions: Use of parallel grafts showed midterm safety and feasibility with low incidence of persistent endoleaks requiring intervention or progression of aneurysm diameter. Reproducible results from other vascular centers will establish the chimney technique in the treatment of type I endoleaks as a viable endovascular modality.

PubMed Disclaimer

Comment in

  • Invited commentary.
    Blankensteijn JD. Blankensteijn JD. J Vasc Surg. 2015 Sep;62(3):584. doi: 10.1016/j.jvs.2015.04.394. J Vasc Surg. 2015. PMID: 26304478 No abstract available.

MeSH terms

LinkOut - more resources