Site-specific tolerance tables and indexing device to improve patient setup reproducibility
- PMID: 26103475
- PMCID: PMC5690111
- DOI: 10.1120/jacmp.v16i3.5097
Site-specific tolerance tables and indexing device to improve patient setup reproducibility
Abstract
While the implementation of tools such as image-guidance and immobilization devices have helped to prevent geometric misses in radiation therapy, many treatments remain prone to error if these items are not available, not utilized for every fraction, or are misused. The purpose of this project is to design a set of site-specific treatment tolerance tables to be applied to the treatment couch for use in a record and verify (R&V) system that will insure accurate patient setup with minimal workflow interruption. This project also called for the construction of a simple indexing device to help insure reproducible patient setup for patients that could not be indexed with existing equipment. The tolerance tables were created by retrospective analysis on a total of 66 patients and 1,308 treatments, separating them into five categories based on disease site: lung, head and neck (H&N), breast, pelvis, and abdomen. Couch parameter tolerance tables were designed to encompass 95% of treatments, and were generated by calculating the standard deviation of couch vertical, longitudinal, and lateral values using the first day of treatment as a baseline. We also investigated an alternative method for generating the couch tolerances by updating the baseline values when patient position was verified with image guidance. This was done in order to adapt the tolerances to any gradual changes in patient setup that would not correspond with a mistreatment. The tolerance tables and customizable indexing device were then implemented for a trial period in order to determine the feasibility of the system. During this trial period we collected data from 1,054 fractions from 65 patients. We then analyzed the number of treatments that would have been out of tolerance, as well as whether or not the tolerances or setup techniques should be adjusted. When the couch baseline values were updated with every imaging fraction, the average rate of tolerance violations was 10% for the lung, H&N, abdomen, and pelvis treatments. Using the indexing device, tolerances for patients with pelvic disease decreased (e.g., from 5.3 cm to 4.3 cm longitudinally). Unfortunately, the results from breast patients were highly variable due to the complexity of the setup technique, making the couch an inadequate surrogate for measuring setup accuracy. In summary, we have developed a method to turn the treatment couch parameters within the R&V system into a useful alert tool, which can be implemented at other institutions, in order to identify potential errors in patient setup.
Figures
Similar articles
-
Eliminating Daily Shifts, Tattoos, and Skin Marks: Streamlining Isocenter Localization With Treatment Plan Embedded Couch Values for External Beam Radiation Therapy.Pract Radiat Oncol. 2019 Jan;9(1):e110-e117. doi: 10.1016/j.prro.2018.08.011. Epub 2018 Oct 21. Pract Radiat Oncol. 2019. PMID: 30355524
-
Assessment of patient setup errors in IGRT in combination with a six degrees of freedom couch.Z Med Phys. 2014 May;24(2):112-22. doi: 10.1016/j.zemedi.2013.11.002. Epub 2014 Jan 10. Z Med Phys. 2014. PMID: 24418323
-
A survey on table tolerances and couch overrides in radiotherapy.J Appl Clin Med Phys. 2016 Nov 8;17(6):405-420. doi: 10.1120/jacmp.v17i6.6261. J Appl Clin Med Phys. 2016. PMID: 27929512 Free PMC article.
-
Analysis of couch position tolerance limits to detect mistakes in patient setup.J Appl Clin Med Phys. 2009 Oct 29;10(4):207-219. doi: 10.1120/jacmp.v10i4.2864. J Appl Clin Med Phys. 2009. PMID: 19918218 Free PMC article.
-
Motion monitoring for cranial frameless stereotactic radiosurgery using video-based three-dimensional optical surface imaging.Med Phys. 2011 Jul;38(7):3981-94. doi: 10.1118/1.3596526. Med Phys. 2011. PMID: 21858995
Cited by
-
A Robust and Affordable Table Indexing Approach for Multi-isocenter Dosimetrically Matched Fields.Cureus. 2017 May 23;9(5):e1270. doi: 10.7759/cureus.1270. Cureus. 2017. PMID: 28652953 Free PMC article.
-
A method to predict patient-specific table coordinates for quality assurance in external beam radiation therapy.J Appl Clin Med Phys. 2018 Sep;19(5):625-631. doi: 10.1002/acm2.12428. Epub 2018 Aug 7. J Appl Clin Med Phys. 2018. PMID: 30085393 Free PMC article.
-
Using a daily monitoring system to reduce treatment position override rates in external beam radiation therapy.J Appl Clin Med Phys. 2022 Jul;23(7):e13629. doi: 10.1002/acm2.13629. Epub 2022 May 4. J Appl Clin Med Phys. 2022. PMID: 35506575 Free PMC article.
-
Automatic couch position calculation using eclipse scripting for external beam radiotherapy.J Appl Clin Med Phys. 2021 Feb;22(2):77-84. doi: 10.1002/acm2.13159. Epub 2021 Jan 13. J Appl Clin Med Phys. 2021. PMID: 33440075 Free PMC article.
-
A self-checking treatment couch coordinate calculation system in radiotherapy.J Appl Clin Med Phys. 2020 Jan;21(1):43-52. doi: 10.1002/acm2.12771. Epub 2019 Nov 18. J Appl Clin Med Phys. 2020. PMID: 31737999 Free PMC article.
References
-
- Ford EC, Terezakis S, Souranis A, Harris K, Gay H, Mutic S. Quality control quantification (QCQ): a tool to measure the value of quality control checks in radiation oncology. Int J Radiat Oncol Biol Phys. 2012;84(3): e263–269. - PubMed
-
- Meeks SL, Bova FJ, Friedman WA, Buatti JM, Moore RD, Mendenhall WM. IRLED‐based patient localization for linac radiosurgery. Int J Radiat Oncol Biol Phys. 1998;41(2):433–39. - PubMed
-
- Peng JL, Kahler D, Li JG, et al. Characterization of a real‐time surface image‐guided stereotactic positioning system. Med Phys. 2010;37(10):5421–33. - PubMed
-
- Yan H, Yin FF, Zhu GP, Ajiouni M, Kim JH. The correlation evaluation of a tumor tracking system using multiple external markers. Med Phys. 2006;33(11):4073–84. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources