Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group
- PMID: 26103482
- PMCID: PMC5690108
- DOI: 10.1120/jacmp.v16i3.5219
Flattening filter-free accelerators: a report from the AAPM Therapy Emerging Technology Assessment Work Group
Abstract
This report describes the current state of flattening filter-free (FFF) radiotherapy beams implemented on conventional linear accelerators, and is aimed primarily at practicing medical physicists. The Therapy Emerging Technology Assessment Work Group of the American Association of Physicists in Medicine (AAPM) formed a writing group to assess FFF technology. The published literature on FFF technology was reviewed, along with technical specifications provided by vendors. Based on this information, supplemented by the clinical experience of the group members, consensus guidelines and recommendations for implementation of FFF technology were developed. Areas in need of further investigation were identified. Removing the flattening filter increases beam intensity, especially near the central axis. Increased intensity reduces treatment time, especially for high-dose stereotactic radiotherapy/radiosurgery (SRT/SRS). Furthermore, removing the flattening filter reduces out-of-field dose and improves beam modeling accuracy. FFF beams are advantageous for small field (e.g., SRS) treatments and are appropriate for intensity-modulated radiotherapy (IMRT). For conventional 3D radiotherapy of large targets, FFF beams may be disadvantageous compared to flattened beams because of the heterogeneity of FFF beam across the target (unless modulation is employed). For any application, the nonflat beam characteristics and substantially higher dose rates require consideration during the commissioning and quality assurance processes relative to flattened beams, and the appropriate clinical use of the technology needs to be identified. Consideration also needs to be given to these unique characteristics when undertaking facility planning. Several areas still warrant further research and development. Recommendations pertinent to FFF technology, including acceptance testing, commissioning, quality assurance, radiation safety, and facility planning, are presented. Examples of clinical applications are provided. Several of the areas in which future research and development are needed are also indicated.
Similar articles
-
IPEM topical report 1: guidance on implementing flattening filter free (FFF) radiotherapy.Phys Med Biol. 2016 Dec 7;61(23):8360-8394. doi: 10.1088/0031-9155/61/23/8360. Epub 2016 Nov 7. Phys Med Biol. 2016. PMID: 27819253
-
MO-A-BRB-02: Facts and Fiction of Flattening Filter Free (FF-FFF) X-Rays Beams.Med Phys. 2012 Jun;39(6Part20):3861-3862. doi: 10.1118/1.4735765. Med Phys. 2012. PMID: 28517518
-
Flattening filter free in intensity-modulated radiotherapy (IMRT) - Theoretical modeling with delivery efficiency analysis.Med Phys. 2019 Jan;46(1):34-44. doi: 10.1002/mp.13267. Epub 2018 Nov 20. Med Phys. 2019. PMID: 30371944
-
Intensity-modulated radiotherapy: current status and issues of interest.Int J Radiat Oncol Biol Phys. 2001 Nov 15;51(4):880-914. doi: 10.1016/s0360-3016(01)01749-7. Int J Radiat Oncol Biol Phys. 2001. PMID: 11704310 Review.
-
Current status and future perspective of flattening filter free photon beams.Med Phys. 2011 Mar;38(3):1280-93. doi: 10.1118/1.3554643. Med Phys. 2011. PMID: 21520840 Review.
Cited by
-
Using FFF Beams to Improve the Therapeutic Ratio of Lung SBRT.J Radiother Pract. 2021 Dec;20(4):419-425. doi: 10.1017/s1460396920000576. Epub 2020 Jul 30. J Radiother Pract. 2021. PMID: 35330584 Free PMC article.
-
FFF-VMAT for SBRT of lung lesions: Improves dose coverage at tumor-lung interface compared to flattened beams.J Appl Clin Med Phys. 2020 Jan;21(1):26-35. doi: 10.1002/acm2.12764. Epub 2019 Dec 20. J Appl Clin Med Phys. 2020. PMID: 31859456 Free PMC article.
-
Monte-Carlo simulation of the Siemens Artiste linear accelerator flat 6 MV and flattening-filter-free 7 MV beam line.PLoS One. 2019 Jan 8;14(1):e0210069. doi: 10.1371/journal.pone.0210069. eCollection 2019. PLoS One. 2019. PMID: 30620742 Free PMC article.
-
Flattening filter free VMAT for a stereotactic, single-dose of 30 Gy to lung lesion in a 15-min treatment slot.J Appl Clin Med Phys. 2020 Apr;21(4):6-12. doi: 10.1002/acm2.12829. Epub 2020 Feb 10. J Appl Clin Med Phys. 2020. PMID: 32039544 Free PMC article.
-
Fast generation of lung SBRT plans with a knowledge-based planning model on ring-mounted Halcyon Linac.J Appl Clin Med Phys. 2021 Nov;22(11):54-63. doi: 10.1002/acm2.13427. Epub 2021 Sep 25. J Appl Clin Med Phys. 2021. PMID: 34562308 Free PMC article.
References
-
- O'Brien PF, Gillies BA, Schwartz M, Young C, Davey P. Radiosurgery with unflattened 6‐MV photon beams. Med Phys. 1991;18(3):519–21. - PubMed
-
- Vassiliev ON, Titt U, Ponisch F, Kry SF, Mohan R, Gillin MT. Dosimetric properties of photon beams from a flattening filter free clinical accelerator. Phys Med Biol. 2006;51(7):1907–17. - PubMed
-
- Cashmore J. The characterization of unflattened photon beams from a 6 MV linear accelerator. Phys Med Biol. 2008;53(7):1933–46. - PubMed
-
- Sixel KE Faddegon BA. Calculation of x‐ray spectra for radiosurgical beams. Med Phys. 1995;22(10):1657–61. - PubMed
-
- Zefkili S, Kappas C, Rosenwald JC. On‐axis and off‐axis primary dose component in high energy photon beams. Med Phys. 1994;21(6):799–808. - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources