Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients
- PMID: 26136992
- PMCID: PMC4473651
- DOI: 10.3892/etm.2015.2412
Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients
Abstract
The aim of the present study was to investigate the X-ray absorption dose and clinical applications of electrocardiogram (ECG)-gated automatic tube current modulation (ATCM) technology in prospective sequential computed tomography coronary angiography (CTCA) scans of overweight patients. A total of 40 patients with a body mass index of >24 were divided at random into groups A and B. Group A consisted of 20 patients, who were examined using ECG-ATCM scanning. For the patients in group A with heart rates <70 bpm, the scanning range was 20-80% of the R-R interval and the patients received full-dose X-rays for 60-80% of the R-R interval. For the group A patients with heart rates >70 bpm, the scanning range was 20-80% of the R-R interval and the patients received full-dose scanning for 35-55% of the R-R interval. For the 20 patients in group B, the scanning range was 20-80% of the R-R interval and patients received a full dose of X-ray radiation for the entire scanning period. The image quality and radiation dose was compared between the two groups. The average radiation dose in groups A and B was 6.91±2.78 and 10.43±3.36, respectively. The radiation doses in group A were reduced by 33.77% when compared with group B (P<0.05). However, there was no marked difference observed in image quality. In summary, using ECG-ATCM technology in prospective sequential CTCA scanning may significantly reduce the radiation dose required for overweight patients.
Keywords: coronary artery; dual-source computed tomography; electrocardiogram automatic tube current modulation technique; low dose; overweight; sequential scanning.
Figures



Similar articles
-
Prospective ECG triggering versus low-dose retrospective ECG-gated 128-channel CT coronary angiography: comparison of image quality and radiation dose.Clin Radiol. 2010 Oct;65(10):809-14. doi: 10.1016/j.crad.2010.05.005. Epub 2010 Jul 21. Clin Radiol. 2010. PMID: 20797467
-
Radiation dose of cardiac dual-source CT: the effect of tailoring the protocol to patient-specific parameters.Eur J Radiol. 2008 Dec;68(3):385-91. doi: 10.1016/j.ejrad.2008.08.015. Epub 2008 Oct 30. Eur J Radiol. 2008. PMID: 18976876
-
[Diagnostic performance of prospective versus retrospective electrocardiogram-gated 64-slice computed tomography coronary angiography].Zhonghua Xin Xue Guan Bing Za Zhi. 2008 Nov;36(11):985-8. Zhonghua Xin Xue Guan Bing Za Zhi. 2008. PMID: 19102910 Chinese.
-
128-slice Dual-source Computed Tomography Coronary Angiography in Patients with Atrial Fibrillation: Image Quality and Radiation Dose of Prospectively Electrocardiogram-triggered Sequential Scan Compared with Retrospectively Electrocardiogram-gated Spiral Scan.Chin Med Sci J. 2013 Jan;27(4):195-200. doi: 10.1016/s1001-9294(13)60001-2. Chin Med Sci J. 2013. PMID: 23294583
-
Intra-individual diagnostic image quality and organ-specific-radiation dose comparison between spiral cCT with iterative image reconstruction and z-axis automated tube current modulation and sequential cCT.Eur J Radiol Open. 2016 Jul 26;3:182-90. doi: 10.1016/j.ejro.2016.05.006. eCollection 2016. Eur J Radiol Open. 2016. PMID: 27504476 Free PMC article. Review.
Cited by
-
Myocardial CT perfusion imaging for the detection of obstructive coronary artery disease: multisegment reconstruction does not improve diagnostic performance.Eur Radiol Exp. 2022 Jan 31;6(1):5. doi: 10.1186/s41747-021-00256-8. Eur Radiol Exp. 2022. PMID: 35099638 Free PMC article. Clinical Trial.
-
Low radiation dose computed tomography coronary angiography: evaluation of the variations in coronary arteries.Surg Radiol Anat. 2016 Dec;38(10):1123-1134. doi: 10.1007/s00276-016-1693-y. Epub 2016 May 18. Surg Radiol Anat. 2016. PMID: 27192982
-
Automatic tube potential selection with tube current modulation in coronary CT angiography: Can it achieve consistent image quality among various individuals?Exp Ther Med. 2018 Jul;16(1):253-259. doi: 10.3892/etm.2018.6158. Epub 2018 May 11. Exp Ther Med. 2018. PMID: 29896246 Free PMC article.
References
LinkOut - more resources
Full Text Sources
Other Literature Sources