. 2015 Jun;9(6):2384-2388.
doi: 10.3892/etm.2015.2412. Epub 2015 Apr 7.

Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients

Affiliations

Dose study of electrocardiogram automatic tube current modulation technology in prospective coronary computed tomography angiography scans of overweight patients

Guiru He et al. Exp Ther Med. 2015 Jun.

Abstract

The aim of the present study was to investigate the X-ray absorption dose and clinical applications of electrocardiogram (ECG)-gated automatic tube current modulation (ATCM) technology in prospective sequential computed tomography coronary angiography (CTCA) scans of overweight patients. A total of 40 patients with a body mass index of >24 were divided at random into groups A and B. Group A consisted of 20 patients, who were examined using ECG-ATCM scanning. For the patients in group A with heart rates <70 bpm, the scanning range was 20-80% of the R-R interval and the patients received full-dose X-rays for 60-80% of the R-R interval. For the group A patients with heart rates >70 bpm, the scanning range was 20-80% of the R-R interval and the patients received full-dose scanning for 35-55% of the R-R interval. For the 20 patients in group B, the scanning range was 20-80% of the R-R interval and patients received a full dose of X-ray radiation for the entire scanning period. The image quality and radiation dose was compared between the two groups. The average radiation dose in groups A and B was 6.91±2.78 and 10.43±3.36, respectively. The radiation doses in group A were reduced by 33.77% when compared with group B (P<0.05). However, there was no marked difference observed in image quality. In summary, using ECG-ATCM technology in prospective sequential CTCA scanning may significantly reduce the radiation dose required for overweight patients.

Keywords: coronary artery; dual-source computed tomography; electrocardiogram automatic tube current modulation technique; low dose; overweight; sequential scanning.

PubMed Disclaimer

Figures

Figure 1.
Figure 1.
For group A patients with heart rates <70 bpm, the scanning range was 20–80% of the R-R interval, with patients receiving full-dose scanning for 60–80% of the R-R interval.
Figure 2.
Figure 2.
Image quality of coronary artery echocardiogram. For group A patients with heart rates >70 bpm, the scanning range was 20–80% of the R-R interval and patients received a full X-ray dose for 35–55% of the R-R interval. Scanning sketch diagram and multiplanar reconstruction map of scanning range of 20–80% interval.
Figure 3.
Figure 3.
Scanning sketch diagram and coronary volume rendered image of group B, with a scanning range 20–80% of the R-R interval.

Similar articles

Cited by

References

    1. Shah DJ, Sachs RK, Wilson DJ. Radiation-induced cancer: a modern view. Br J Radiol. 2012;85:e1166–e1173. doi: 10.1259/bjr/25026140. - DOI - PMC - PubMed
    1. Sodickson A, Baeyens PF, Andriole KP, et al. Recurrent CT, cumulative radiation exposure, and associated radiation-induced cancer risks from CT of adults. Radiology. 2009;251:175–184. doi: 10.1148/radiol.2511081296. - DOI - PubMed
    1. Husmann L, Valeta I, Gaemperli O, et al. Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J. 2008;29:191–197. doi: 10.1093/eurheartj/ehm613. - DOI - PubMed
    1. Buechel RR, Husmann L, Herzog BA, et al. Low-dose computed tomography coronary angiography with prospective electrocardiogram triggering: feasibility in a large population. J Am Coll Cardiol. 2011;57:332–336. doi: 10.1016/j.jacc.2010.08.634. - DOI - PubMed
    1. Achenbach S, Goroll T, Seltmann M, et al. Detection of coronary artery stenoses by low-dose, prospectively ECG-triggered, high-pitch spiral coronary CT angiography. JACC Cardiovasc Imaging. 2011;4:328–337. doi: 10.1016/j.jcmg.2011.01.012. - DOI - PubMed

LinkOut - more resources