Risk of Misdiagnosis Due to Allele Dropout and False-Positive PCR Artifacts in Molecular Diagnostics: Analysis of 30,769 Genotypes

Affiliations
Free article

Risk of Misdiagnosis Due to Allele Dropout and False-Positive PCR Artifacts in Molecular Diagnostics: Analysis of 30,769 Genotypes

Jonatan Blais et al. J Mol Diagn. 2015 Sep.
Free article

Abstract

Quality control is a complex issue for clinical molecular diagnostic applications. In the case of genotyping assays, artifacts such as allele dropout represent a risk of misdiagnosis for amplification-based methods. However, its frequency of occurrence in PCR-based diagnostic assays remains unknown. To maximize the likelihood of detecting allele dropout, our clinical genotyping PCR-based assays are designed with two independent assays for each allele (nonoverlapping primers on each DNA strand). To estimate the incidence of allelic dropout, we took advantage of the capacity of our clinical assays to detect such events. We retrospectively studied their occurrence in the initial PCR assay for 30,769 patient reports for mutations involved in four diseases produced over 8 years. Ninety-three allele dropout events were detected and all were solved before reporting. In addition, 42 cases of artifacts caused by amplification of an allele ultimately confirmed to not be part of the genotype (drop-in events) were detected and solved. These artifacts affected 1:227 genotypes, 94% of which were due to nonreproducible PCR failures rather than sequence variants interfering with the assay, suggesting that careful primer design cannot prevent most of these errors. This provides a quantitative estimate for clinical laboratories to take this phenomenon into account in quality management and to favor assay designs that can detect (and minimize) occurrence of these artifacts in routine clinical use.