In vivo measurements of bone lead--a comparison of two x-ray fluorescence techniques used at three different bone sites
- PMID: 2616639
- DOI: 10.1088/0031-9155/34/12/007
In vivo measurements of bone lead--a comparison of two x-ray fluorescence techniques used at three different bone sites
Abstract
In vivo bone lead measurements have been made on a group of about 120 people, most of whom were lead exposed workers. Two different x-ray fluorescence (XRF) techniques were used to make measurements at three bone sites. Finger lead was measured using 57Co sources, and lead measurements were made in both tibia and calcaneus with a technique based on 109Cd sources. The results of the bone lead measurements correlated strongly with each other and with the index of cumulative exposure, thus confirming the value and reliability of these in vivo measurements as a tool in the study of chronic lead exposure. Measurement precision, +/- 1 standard deviation, was highest for tibia +/- 7.4 micrograms (g bone mineral)-1, +/- 16.6 micrograms (g bone mineral)-1 for the calcaneus and lowest for phalangeal lead +/- 25.0 micrograms (g bone mineral)-1. Maximum absorbed doses to the skin were comparable for all three measurements (1-3 mGy). The mean whole body dose equivalents were all low, but that for the finger measurement, 0.1 microSv, was significantly less than for the calcaneus and tibia measurements 3-5 microSv.
Similar articles
-
In vivo X-ray fluorescence of lead in bone: review and current issues.Environ Health Perspect. 1994 Feb;102(2):172-7. doi: 10.1289/ehp.94102172. Environ Health Perspect. 1994. PMID: 8033846 Free PMC article. Review.
-
In vivo measurement of lead in bone using x-ray fluorescence.Phys Med Biol. 1985 Sep;30(9):929-43. doi: 10.1088/0031-9155/30/9/005. Phys Med Biol. 1985. PMID: 4048276
-
In vivo measurements of lead in bone in long-term exposed lead smelter workers.Arch Environ Health. 1993 May-Jun;48(3):147-56. doi: 10.1080/00039896.1993.9940813. Arch Environ Health. 1993. PMID: 8333784
-
Factors affecting in vivo measurement precision and accuracy of 109Cd K x-ray fluorescence measurements.Phys Med Biol. 1999 Sep;44(9):2263-73. doi: 10.1088/0031-9155/44/9/313. Phys Med Biol. 1999. PMID: 10495120
-
Lead in bone: sampling and quantitation using K X-rays excited by 109Cd.Environ Health Perspect. 1991 Feb;91:49-55. doi: 10.1289/ehp.919149. Environ Health Perspect. 1991. PMID: 2040251 Free PMC article. Review.
Cited by
-
Assessment of lead exposure risk in locksmiths.Int J Environ Res Public Health. 2005 Apr;2(1):164-9. doi: 10.3390/ijerph2005010164. Int J Environ Res Public Health. 2005. PMID: 16705814 Free PMC article.
-
XRF-measured bone lead (Pb) as a biomarker for Pb exposure and toxicity among children diagnosed with Pb poisoning.Biomarkers. 2016;21(4):347-52. doi: 10.3109/1354750X.2016.1139183. Epub 2016 Feb 9. Biomarkers. 2016. PMID: 26856822 Free PMC article.
-
In vivo X-ray fluorescence of lead in bone: review and current issues.Environ Health Perspect. 1994 Feb;102(2):172-7. doi: 10.1289/ehp.94102172. Environ Health Perspect. 1994. PMID: 8033846 Free PMC article. Review.
-
In vivo measurements of lead in bone at four anatomical sites: long term occupational and consequent endogenous exposure.Br J Ind Med. 1992 Sep;49(9):631-44. doi: 10.1136/oem.49.9.631. Br J Ind Med. 1992. PMID: 1390269 Free PMC article.
-
The reproducibility of 109Cd-based X-ray fluorescence measurements of bone lead.Environ Health Perspect. 1994 Aug;102(8):690-4. doi: 10.1289/ehp.94102690. Environ Health Perspect. 1994. PMID: 7895710 Free PMC article.