Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Review
. 2015 Jul-Sep;22(3):265-70.
doi: 10.4103/0974-9233.159691.

Optical Issues in Measuring Strabismus

Affiliations
Review

Optical Issues in Measuring Strabismus

Kristina Irsch. Middle East Afr J Ophthalmol. 2015 Jul-Sep.

Abstract

Potential errors and complications during examination and treatment of strabismic patients can be reduced by recognition of certain optical issues. This articles reviews basic as well as guiding principles of prism optics and optics of the eye to equip the reader with the necessary know-how to avoid pitfalls that are commonly encountered when using prisms to measure ocular deviations (e.g., during cover testing), and when observing the corneal light reflex to estimate ocular deviations (e.g., during Hirschberg or Krimsky testing in patients who do not allow for cover testing using prisms).

Keywords: Angle kappa; Prentice position; Prentice's rule; angle lambda; frontal plane position; minimum deviation position; prism diopters.

PubMed Disclaimer

Conflict of interest statement

Conflict of Interest: None declared.

Figures

Figure 1
Figure 1
Light path through a 15Δ prism. The amount of deflection (in cm) of a light ray that is caused by a prism, measured 1 m (100 cm) away from the prism, is equal to the power of the prism in prism diopters (Δ). Thus, a 15Δ prism, as in this illustration, deviates a light ray 15 cm toward the base of the prism, when measured 1 m away from it
Figure 2
Figure 2
Relationship between prism diopters and degrees. For angles smaller than 45° (or 100Δ), the number of prism diopters per degree is about 2. For angles larger than 45° (or 100Δ), this approximation becomes invalid, and as one approaches 90°, the number of prism diopters per degree goes up to infinity
Figure 3
Figure 3
Positioning of prisms. (a) Prentice position. Glass prisms are calibrated for use in this position, so the line of sight makes a right angle with one of the surfaces. (b) Minimum deviation position. Plastic prisms, including plastic prism bars, are calibrated in this position, so the line of sight makes an equal angle with each prism surface. (c) Frontal plane position. Holding plastic prisms in this position, with the back surface flat to the face of the patient, closely approximates the minimum deviation position, which would otherwise be difficult to estimate in clinical practice
Figure 4
Figure 4
Interface between two stacked prisms. While the first glass prism is in the Prentice position, with the light ray being perpendicular to the first surface of the prism, the second glass prism is nowhere near the Prentice position, with the light ray going in at an angle far from perpendicular (adapted from reference 4)
Figure 5
Figure 5
Detection of prism with lensmeter. Displacement of the cross-line target away from the reticle center indicates presence of prism, with the decentration in the direction of the prism base. In this illustration, the cross-line target is displaced 4 rings to the right, indicating the presence of 2Δ base-in or-out prism if dealing with a spectacle lens for the right eye or left eye respectively
Figure 6
Figure 6
Schematic representation of angle lambda and corneal light reflex positioning. (a) Normal angle lambda (λ) between the patient's line of sight (directed at the examiner's handlight) and pupillary axis is associated with slight nasalward positioning of the corneal light reflex (the reflection of the handlight from the anterior surface of the cornea) with respect to the pupillary center. (b) Eye with normal angle lambda (λ) is not fixating on the handlight, with strabismic deviation (δ), showing nasalward displacement of the corneal light reflex from the pupillary center. Note the imaginary string that connects the corneal center of curvature (CR) with the handlight. (c) Large angle lambda (λ+) caused by a temporally dragged, but fixating fovea, showing the same amount of nasalward displacement of the corneal light reflex from the pupillary center as in (b) with the deviated eye with normal angle lambda (λ) not fixating on the handlight

References

    1. Guyton DL, West CE, Miller JM, Wisnicki HJ. Ophthalmic Optics and Clinical Refraction. Baltimore: Prism Press; 1999.
    1. Guyton DL, Miller JM, West CE. Optical pearls and pitfalls. In: Wright KW, Spiegel PH, editors. Pediatric Ophthalmology and Strabismus. New York: Springer-Verlag; 2003.
    1. Gutmark R, Guyton DL, Irsch K. Prescribing prisms. Focal points: Clinical modules for ophthalmologists. Am Acad Ophthalmol. 2011;XXIX:10. (Module 1 of 3); Available from: http://store.aao.org/focal-points-2011-module-prescribing-prisms.html .
    1. Thompson JT, Guyton DL. Ophthalmic prisms. Measurement errors and how to minimize them. Ophthalmology. 1983;90:204–10. - PubMed
    1. Thompson JT, Guyton DL. Ophthalmic prisms. Deviant behavior at near. Ophthalmology. 1985;92:684–90. - PubMed