Point-of-care testing and INR within-subject variation in patients receiving a constant dose of vitamin K antagonist
- PMID: 26202616
- DOI: 10.1160/TH15-02-0128
Point-of-care testing and INR within-subject variation in patients receiving a constant dose of vitamin K antagonist
Abstract
Many patients treated with vitamin K antagonists (VKA) determine their INR using point-of-care (POC) whole blood coagulation monitors. The primary aim of the present study was to assess the INR within-subject variation in self-testing patients receiving a constant dose of VKA. The second aim of the study was to derive INR imprecision goals for whole blood coagulation monitors. Analytical performance goals for INR measurement can be derived from the average biological within-subject variation. Fifty-six Thrombosis Centres in the Netherlands were invited to select self-testing patients who were receiving a constant dose of either acenocoumarol or phenprocoumon for at least six consecutive INR measurements. In each patient, the coefficient of variation (CV) of INRs was calculated. One Thrombosis Centre selected regular patients being monitored with a POC device by professional staff. Sixteen Dutch Thrombosis Centres provided results for 322 selected patients, all using the CoaguChek XS. The median within-subject CV in patients receiving acenocoumarol (10.2 %) was significantly higher than the median CV in patients receiving phenprocoumon (8.6 %) (p = 0.001). The median CV in low-target intensity acenocoumarol self-testing patients (10.4 %) was similar to the median CV in regular patients monitored by professional staff (10.2 %). Desirable INR analytical imprecision goals for POC monitoring with CoaguChek XS in patients receiving either low-target intensity acenocoumarol or phenprocoumon were 5.1 % and 4.3 %, respectively. The approximate average value for the imprecision of the CoaguChek XS, i. e. 4 %, is in agreement with these goals.
Keywords: International Normalized Ratio; Vitamin K–antagonist; analytical performance goal; point-of-care testing; within-subject variation.
Similar articles
-
Variability of INR in patients on stable long-term treatment with phenprocoumon and acenocoumarol and implications for analytical quality requirements.Thromb Haemost. 2009 Sep;102(3):588-92. doi: 10.1160/TH09-02-0092. Thromb Haemost. 2009. PMID: 19718481
-
International normalized ratio monitoring of vitamin K antagonist therapy: comparative performance of point-of-care and laboratory-derived testing.Semin Thromb Hemost. 2015 Apr;41(3):279-86. doi: 10.1055/s-0035-1549091. Epub 2015 Apr 3. Semin Thromb Hemost. 2015. PMID: 25839866
-
Accuracy and clinical usefulness of the CoaguChek S and XS Point of Care devices when starting warfarin in a hospital outreach setting.Thromb Res. 2009 Apr;123(6):909-13. doi: 10.1016/j.thromres.2008.10.006. Epub 2008 Nov 29. Thromb Res. 2009. PMID: 19042011
-
[Near-patient testing devices to monitor vitamin K antagonists].Ann Biol Clin (Paris). 2013 Nov;71:59-69. doi: 10.1684/abc.2013.0904. Ann Biol Clin (Paris). 2013. PMID: 24235329 Review. French.
-
Approaches to optimal dosing of vitamin K antagonists.Semin Thromb Hemost. 2012 Oct;38(7):667-72. doi: 10.1055/s-0032-1324713. Epub 2012 Aug 14. Semin Thromb Hemost. 2012. PMID: 22893509 Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical