Toric Intraocular Lens Calculations Using Ratio of Anterior to Posterior Corneal Cylinder Power
- PMID: 26215437
- DOI: 10.1016/j.ajo.2015.07.011
Toric Intraocular Lens Calculations Using Ratio of Anterior to Posterior Corneal Cylinder Power
Abstract
Purpose: To evaluate the accuracy of toric intraocular lens (IOL) calculation using estimated total corneal astigmatism based on the anterior-to-posterior corneal cylinder power ratio according to the axis orientation of anterior corneal astigmatism.
Design: Retrospective cross-sectional study.
Methods: Nine hundred twenty-eight eyes of 928 reference subjects and 20 cataract patients (20 eyes) implanted with a toric IOL were enrolled. Linear regression analysis parameters (β0 and β1) of relationship between the simulated keratometry cylinder (CylSimK) and posterior corneal cylinder power of reference subjects were used to calculate the estimated posterior corneal astigmatism (-[β1 × CylSimK + β0] @ 90). When regression analysis was not significant, estimated posterior corneal astigmatism was defined as the negative value of the mean posterior corneal cylinder power @ 90. Estimated total corneal astigmatism was defined as the vectorial sum of anterior corneal astigmatism and estimated posterior corneal astigmatism. Residual astigmatism error, predicted using SimK, was compared with that predicted using estimated total corneal astigmatism.
Results: Estimated posterior corneal astigmatism was determined to be -(0.15 × CylSimK + 0.22) @ 90 in eyes with with-the-rule astigmatism, -(0.05 × CylSimK + 0.27) @ 90 in oblique astigmatism, and -0.27 @ 90 in against-the-rule astigmatism. The median magnitude of the predicted residual astigmatism error calculated using estimated total corneal astigmatism (0.30 cylinder diopters) was significantly smaller than that calculated with SimK (0.50 cylinder diopters).
Conclusions: Toric IOL calculations using estimated total corneal astigmatism based on the anterior-to-posterior corneal cylinder power ratio provided a more appropriate toric IOL cylinder power than calculations using SimK astigmatism.
Copyright © 2015 Elsevier Inc. All rights reserved.
Similar articles
-
Comparison of astigmatism prediction error taken with the Pentacam measurements, Baylor nomogram, and Barrett formula for toric intraocular lens implantation.BMC Ophthalmol. 2017 Aug 24;17(1):156. doi: 10.1186/s12886-017-0550-z. BMC Ophthalmol. 2017. PMID: 28836956 Free PMC article.
-
Effect of posterior corneal astigmatism on power calculation and alignment of toric intraocular lenses: Comparison of methodologies.J Cataract Refract Surg. 2016 Feb;42(2):217-25. doi: 10.1016/j.jcrs.2015.11.036. J Cataract Refract Surg. 2016. PMID: 27026445
-
Impact of Posterior Corneal Surface on Toric Intraocular Lens (IOL) Calculation.Curr Eye Res. 2015;40(8):809-14. doi: 10.3109/02713683.2014.959708. Epub 2014 Sep 26. Curr Eye Res. 2015. PMID: 25259550
-
A new slant on toric intraocular lens power calculation.J Refract Surg. 2013 May;29(5):348-54. doi: 10.3928/1081597X-20130415-06. J Refract Surg. 2013. PMID: 23659233 Review.
-
The role of posterior corneal power in 21st century biometry: A review.J Fr Ophtalmol. 2021 Sep;44(7):1052-1058. doi: 10.1016/j.jfo.2020.11.031. Epub 2021 Jun 17. J Fr Ophtalmol. 2021. PMID: 34148699 Review.
Cited by
-
Comparison of Refractive Prediction Errors of Toric Intraocular Lens Formulae.Cureus. 2025 Feb 26;17(2):e79679. doi: 10.7759/cureus.79679. eCollection 2025 Feb. Cureus. 2025. PMID: 40161103 Free PMC article.
-
The Influence of Analysis Mode Selection on Prediction Accuracy of Corneal Astigmatism Using Pentacam.Front Med (Lausanne). 2021 Oct 15;8:713502. doi: 10.3389/fmed.2021.713502. eCollection 2021. Front Med (Lausanne). 2021. PMID: 34722562 Free PMC article.
-
Image-guided lens extraction surgery: a systematic review.Int J Ophthalmol. 2019 Jan 18;12(1):135-151. doi: 10.18240/ijo.2019.01.21. eCollection 2019. Int J Ophthalmol. 2019. PMID: 30662853 Free PMC article. Review.
-
Clinical outcomes with toric intraocular lenses planned using an optical low coherence reflectometry ocular biometer with a new toric calculator.Clin Ophthalmol. 2016 Nov 3;10:2141-2147. doi: 10.2147/OPTH.S120414. eCollection 2016. Clin Ophthalmol. 2016. PMID: 27843285 Free PMC article.
-
Agreement between total corneal astigmatism calculated by vector summation and total corneal astigmatism measured by ray tracing using Galilei double Scheimpflug analyzer.J Optom. 2018 Apr-Jun;11(2):113-120. doi: 10.1016/j.optom.2017.05.003. Epub 2017 Aug 4. J Optom. 2018. PMID: 28780995 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical