Biplanar Open Wedge High Tibial Osteotomy in the Medial Compartment Osteoarthritis of the Knee Joint: Comparison between the Aescula and TomoFix Plate
- PMID: 26217464
- PMCID: PMC4515458
- DOI: 10.4055/cios.2015.7.2.185
Biplanar Open Wedge High Tibial Osteotomy in the Medial Compartment Osteoarthritis of the Knee Joint: Comparison between the Aescula and TomoFix Plate
Abstract
Background: The purpose of this study was to compare the results of Aescula and TomoFix plates used for biplanar open wedge high tibial osteotomy in medial osteoarthritis of the knee joint with varus deformity.
Methods: A consecutive series of 50 cases of biplanar open wedge high tibial osteotomy were evaluated retrospectively. Group A contained 25 cases treated by using the Aescula plate, and group T contained 25 cases treated by using the TomoFix plate. Full weight-bearing was permitted at 6 weeks after surgery in group A and at 2 weeks in group T. Clinical evaluations were performed at the final follow-up by using postoperative knee scores and functional scores. Radiographic analysis included postoperative mechanical femur-tibia angle, change in posterior tibial slope angle, and complications related to implants. The mean follow-up periods were 30 months in group A and 26 months in group T.
Results: The knee and functional scores were improved at the final follow-up in both groups (p < 0.05), but no differences were observed between the two groups (p > 0.05). An acceptable correction angle was obtained in 52% of group A and in 84% of group T (p = 0.015). Change in posterior tibial slope angle was larger in group A than in group T (p < 0.001), showing better maintenance of posterior tibial slope in group T. In group A, there were 3 cases of screw loosening and 4 cases of delayed union. In addition, there were residual varus deformities in 7 cases (6 in group A and 1 in group T).
Conclusions: This study shows that firm fixation using a TomoFix plate for open wedge high tibial osteotomy produces better radiologic results and a low complication rate than those of the Aescula spacer plate.
Keywords: Aescula plate; Comparative study; Knee; Osteoarthritis; Osteotomy; TomoFix plate.
Conflict of interest statement
Figures



Similar articles
-
The short-term follow-up results of open wedge high tibial osteotomy with using an Aescula open wedge plate and an allogenic bone graft: the minimum 1-year follow-up results.Clin Orthop Surg. 2010 Mar;2(1):47-54. doi: 10.4055/cios.2010.2.1.47. Epub 2010 Feb 4. Clin Orthop Surg. 2010. PMID: 20191001 Free PMC article.
-
Comparative study of medial opening-wedge high tibial osteotomy using 2 different implants.Arthroscopy. 2013 Jun;29(6):1063-71. doi: 10.1016/j.arthro.2013.02.020. Epub 2013 Apr 26. Arthroscopy. 2013. PMID: 23623294
-
High complication rate after biplanar open wedge high tibial osteotomy stabilized with a new spacer plate (Position HTO plate) without bone substitute.Arthroscopy. 2011 May;27(5):644-52. doi: 10.1016/j.arthro.2011.01.008. Arthroscopy. 2011. PMID: 21663721 Clinical Trial.
-
Accuracy of frontal and sagittal plane correction in open-wedge high tibial osteotomy.Arthroscopy. 2004 Apr;20(4):366-72. doi: 10.1016/j.arthro.2004.01.024. Arthroscopy. 2004. PMID: 15067275 Review.
-
Opening- and closing-wedge high tibial osteotomy are comparable and early full weight bearing is safe with angular stable plate fixation: a meta-analysis.Knee Surg Sports Traumatol Arthrosc. 2023 Jul;31(7):3015-3026. doi: 10.1007/s00167-022-07229-3. Epub 2022 Dec 7. Knee Surg Sports Traumatol Arthrosc. 2023. PMID: 36473985 Review.
Cited by
-
Change in adduction moment following medial open wedge high tibial osteotomy: a meta-analysis.BMC Musculoskelet Disord. 2019 Mar 6;20(1):102. doi: 10.1186/s12891-019-2472-9. BMC Musculoskelet Disord. 2019. PMID: 30841871 Free PMC article.
-
High Tibial Osteotomy for Genu Varum in Adults: Do Proprietary Implants Limit the Quality of Correction?Strategies Trauma Limb Reconstr. 2020 Jan-Apr;15(1):13-22. doi: 10.5005/jp-journals-10080-1449. Strategies Trauma Limb Reconstr. 2020. PMID: 33363636 Free PMC article.
-
Comparison of Anatomical Conformity between TomoFix Anatomical Plate and TomoFix Conventional Plate in Open-Wedge High Tibial Osteotomy.Medicina (Kaunas). 2022 Aug 3;58(8):1045. doi: 10.3390/medicina58081045. Medicina (Kaunas). 2022. PMID: 36013511 Free PMC article.
-
Effect of the sagittal osteotomy inclination angle on the posterior tibial slope change in high tibial osteotomy: three-dimensional simulation study.Sci Rep. 2022 Nov 10;12(1):19254. doi: 10.1038/s41598-022-23412-5. Sci Rep. 2022. PMID: 36357467 Free PMC article.
-
Static and fatigue strength of a novel anatomically contoured implant compared to five current open-wedge high tibial osteotomy plates.J Exp Orthop. 2017 Dec 8;4(1):39. doi: 10.1186/s40634-017-0115-3. J Exp Orthop. 2017. PMID: 29222607 Free PMC article.
References
-
- Hofmann AA, Kane KR. Total knee arthroplasty after high tibial osteotomy. Orthopedics. 1994;17(9):887–890. - PubMed
-
- Keene JS, Dyreby JR., Jr High tibial osteotomy in the treatment of osteoarthritis of the knee: the role of preoperative arthroscopy. J Bone Joint Surg Am. 1983;65(1):36–42. - PubMed
-
- Pape D, Kohn D, van Giffen N, Hoffmann A, Seil R, Lorbach O. Differences in fixation stability between spacer plate and plate fixator following high tibial osteotomy. Knee Surg Sports Traumatol Arthrosc. 2013;21(1):82–89. - PubMed
-
- Brinkman JM, Lobenhoffer P, Agneskirchner JD, Staubli AE, Wymenga AB, van Heerwaarden RJ. Osteotomies around the knee: patient selection, stability of fixation and bone healing in high tibial osteotomies. J Bone Joint Surg Br. 2008;90(12):1548–1557. - PubMed
-
- van den Bekerom MP, Patt TW, Kleinhout MY, van der Vis HM, Albers GH. Early complications after high tibial osteotomy: a comparison of two techniques. J Knee Surg. 2008;21(1):68–74. - PubMed