Comparison of interlaboratory results for blood lead with results from a definitive method
- PMID: 262177
Comparison of interlaboratory results for blood lead with results from a definitive method
Abstract
Results reported by 113 participants in the Blood Lead Proficiency Testing Program conducted by the Center for Disease Control were compared with those obtained by the National Bureau of Standards (NBS) with a definitive methods (mass spectroscopy-isotopic dilution) for blood lead analyses. Data were compiled from the results obtained for 12 whole-blood samples containing 1.5 g of disodium EDTA per liter. Twelve separate blood samples were obtained from cattle which had been given lead nitrate orally. Lead concentrations in the samples ranged from 0.628 to 4.93 mumol/L (130-1020 micrograms/L) as determined by NBS. The methods used by laboratories were classified according to six basic groups: anodic stripping voltametry; and atomic absorption spectroscopy in which either extraction, carbon rod, graphite furnace, tantalum strip, or Delves cup was used. For results obtained in each group a linear regression analyses of laboratory values was made on the basis of NBS values. In comparison to the definitive method, most field methods for blood lead tended to overestimate the lead concentration when the actual lead concentration was less than 1.96 mumol/L (400 micrograms/L) and to underestimate the lead concentration when the actual lead concentration was greater than 2.45 mumol/L (500 micrograms/L).
Similar articles
-
Graphite furnace atomic absorption spectroscopic measurement of blood lead in matrix-matched standards.Clin Chem. 1994 Sep;40(9):1730-4. Clin Chem. 1994. PMID: 8070083
-
Disparities in clinical laboratory performance for blood lead analysis.Arch Pediatr Adolesc Med. 1996 Jun;150(6):609-14. doi: 10.1001/archpedi.1996.02170310043008. Arch Pediatr Adolesc Med. 1996. PMID: 8646311
-
Monitoring human exposure to lead: an assessment of current laboratory performance for the determination of blood lead.Environ Res. 1992 Apr;57(2):149-62. doi: 10.1016/s0013-9351(05)80075-1. Environ Res. 1992. PMID: 1568437
-
Blood lead proficiency testing: overview of the federally sponsored program in the U.S.J Int Fed Clin Chem. 1993 Sep;5(4):158-61. J Int Fed Clin Chem. 1993. PMID: 10146391 Review.
-
Blood lead levels in children, China.Environ Res. 2006 Jul;101(3):412-8. doi: 10.1016/j.envres.2005.11.007. Epub 2006 Jan 25. Environ Res. 2006. PMID: 16442092 Review.
Cited by
-
Lead absorption in indoor firing range users.Am J Public Health. 1989 Aug;79(8):1029-32. doi: 10.2105/ajph.79.8.1029. Am J Public Health. 1989. PMID: 2751019 Free PMC article.
-
Lead poisoning from retained bullets. Pathogenesis, diagnosis, and management.Ann Surg. 1982 Mar;195(3):305-13. doi: 10.1097/00000658-198203000-00010. Ann Surg. 1982. PMID: 6800314 Free PMC article.
-
Measurement Challenges at Low Blood Lead Levels.Pediatrics. 2017 Aug;140(2):e20170272. doi: 10.1542/peds.2017-0272. Epub 2017 Jul 17. Pediatrics. 2017. PMID: 28771411 Free PMC article.
-
Hemolyzed, lyophilized bovine blood for quality control of lead determination of human whole blood.Bull Environ Contam Toxicol. 1985 Sep;35(3):380-5. doi: 10.1007/BF01636526. Bull Environ Contam Toxicol. 1985. PMID: 4041649 No abstract available.
-
Storage and preservation of blood and urine for trace element analysis. A review.Biol Trace Elem Res. 1995 Aug-Sep;49(2-3):187-210. doi: 10.1007/BF02788968. Biol Trace Elem Res. 1995. PMID: 8562287 Review.
MeSH terms
Substances
LinkOut - more resources
Research Materials
Miscellaneous