Adverse local tissue reactions in metal-on-polyethylene total hip arthroplasty due to trunnion corrosion: the risk of misdiagnosis
- PMID: 26224816
- DOI: 10.1302/0301-620X.97B8.34682
Adverse local tissue reactions in metal-on-polyethylene total hip arthroplasty due to trunnion corrosion: the risk of misdiagnosis
Erratum in
-
Errata/Corrigenda.Bone Joint J. 2015 Dec;97-B(12):1732. doi: 10.1302/0301-620X.97B7.37439e. Bone Joint J. 2015. PMID: 26637692 Free PMC article.
Abstract
Adverse reaction to wear and corrosion debris is a cause for concern in total hip arthroplasty (THA). Modular junctions are a potential source of such wear products and are associated with secondary pseudotumour formation. We present a consecutive series of 17 patients treated at our unit for this complication following metal-on-highly cross-linked polyethylene (MoP) THA. We emphasise the risk of misdiagnosis as infection, and present the aggregate laboratory results and pathological findings in this series. The clinical presentation was pain, swelling or instability. Solid, cystic and mixed soft-tissue lesions were noted on imaging and confirmed intra-operatively. Corrosion at the head-neck junction was noted in all cases. No bacteria were isolated on multiple pre- and intra-operative samples yet the mean erythrocyte sedimentation rate was 49 (9 to 100) and C-reactive protein 32 (0.6 to 106) and stromal polymorphonuclear cell counts were noted in nine cases. Adverse soft-tissue reactions can occur in MoP THA owing to corrosion products released from the head-neck junction. The diagnosis should be carefully considered when investigating pain after THA. This may avoid the misdiagnosis of periprosthetic infection with an unidentified organism and mitigate the unnecessary management of these cases with complete single- or two-stage exchange.
Keywords: Arthroplasty; Hip Replacement; Complication; Post-operative; Diagnosis.
©2015 The British Editorial Society of Bone & Joint Surgery.
Similar articles
-
Adverse local tissue reaction arising from corrosion at the femoral neck-body junction in a dual-taper stem with a cobalt-chromium modular neck.J Bone Joint Surg Am. 2013 May 15;95(10):865-72. doi: 10.2106/JBJS.L.01042. J Bone Joint Surg Am. 2013. PMID: 23677352 Free PMC article.
-
Diagnosis and Management of Adverse Local Tissue Reactions Secondary to Corrosion at the Head-Neck Junction in Patients With Metal on Polyethylene Bearings.J Arthroplasty. 2016 Jan;31(1):264-8. doi: 10.1016/j.arth.2015.07.039. Epub 2015 Aug 1. J Arthroplasty. 2016. PMID: 26321628
-
Revision for adverse local tissue reaction following metal-on-polyethylene total hip arthroplasty is associated with a high risk of early major complications.Bone Joint J. 2018 Jun 1;100-B(6):720-724. doi: 10.1302/0301-620X.100B6.BJJ-2017-1466.R1. Bone Joint J. 2018. PMID: 29855241
-
The local effects of metal corrosion in total hip arthroplasty.Orthop Clin North Am. 2014 Jan;45(1):9-18. doi: 10.1016/j.ocl.2013.08.003. Epub 2013 Sep 21. Orthop Clin North Am. 2014. PMID: 24267203 Review.
-
The surgical options and clinical evidence for treatment of wear or corrosion occurring with THA or TKA.Clin Orthop Relat Res. 2014 Dec;472(12):3674-86. doi: 10.1007/s11999-014-3652-4. Clin Orthop Relat Res. 2014. PMID: 25024023 Free PMC article. Review.
Cited by
-
Retrieval analysis of metal and ceramic femoral heads on a single CoCr stem design.Bone Joint Res. 2017 May;6(5):345-350. doi: 10.1302/2046-3758.65.BJR-2016-0325.R1. Bone Joint Res. 2017. PMID: 28566326 Free PMC article.
-
Do Polyethylene Supra-Macroparticles Lead to Pseudotumor Formation in Metal-on-Polyethylene Total Hip Arthroplasty?Arthroplast Today. 2020 Jul 23;6(3):526-531. doi: 10.1016/j.artd.2020.06.006. eCollection 2020 Sep. Arthroplast Today. 2020. PMID: 32743035 Free PMC article.
-
Advanced MRI Techniques for the Hip Joint: Focus on the Postoperative Hip.AJR Am J Roentgenol. 2017 Sep;209(3):534-543. doi: 10.2214/AJR.16.17789. Epub 2017 May 24. AJR Am J Roentgenol. 2017. PMID: 28537800 Free PMC article. Review.
-
The risk of developing cancer following metal-on-metal hip replacement compared with non metal-on-metal hip bearings: Findings from a prospective national registry "The National Joint Registry of England, Wales, Northern Ireland and the Isle of Man".PLoS One. 2018 Sep 20;13(9):e0204356. doi: 10.1371/journal.pone.0204356. eCollection 2018. PLoS One. 2018. PMID: 30235326 Free PMC article.
-
Delay of total hip arthroplasty to advanced stage worsens post-operative hip motion in patients with femoral head osteonecrosis.Int Orthop. 2018 Jul;42(7):1599-1603. doi: 10.1007/s00264-018-3952-5. Epub 2018 Apr 26. Int Orthop. 2018. PMID: 29700582
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials