The impact of pediatric-specific dose modulation curves on radiation dose and image quality in head computed tomography
- PMID: 26242810
- DOI: 10.1007/s00247-015-3398-5
The impact of pediatric-specific dose modulation curves on radiation dose and image quality in head computed tomography
Abstract
Background: The volume of CT examinations has increased with resultant increases in collective dose values over the last decade.
Objective: To analyze the impact of the tube current and voltage modulation for dose values and image quality of pediatric head CT examinations.
Materials and methods: Head CT examinations were performed on anthropomorphic phantoms and four pediatric age categories before and after the introduction of dedicated pediatric curves for tube voltage and current modulation. Local diagnostic reference levels were calculated. Visual grading characteristic image quality evaluation was performed by four pediatric neuroradiologists and image noise comparisons were performed.
Results: Pediatric-specific modulation curves demonstrated a 49% decrease in mean radiation dose for phantom examinations. The local diagnostic reference levels (CTDIvol) for clinical examinations decreased by 52%, 41%, 46% and 40% for newborn, 5-, 10- and 15-year-old patients, respectively. Visual grading characteristic image quality was maintained for the majority of age categorizations (area under the curve = 0.5) and image noise measurements did not change (P = 0.693).
Conclusion: Pediatric-specific dose modulation curves resulted in an overall mean dose reduction of 45% with no significant differences in subjective or objective image quality findings.
Keywords: Child; Computed tomography; Head; Radiation dose reduction.
Similar articles
-
Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography: A Phantom Study on Radiation Dose and Image Quality.Invest Radiol. 2019 May;54(5):265-272. doi: 10.1097/RLI.0000000000000537. Invest Radiol. 2019. PMID: 30562273 Free PMC article.
-
Ultra-high pitch chest computed tomography at 70 kVp tube voltage in an anthropomorphic pediatric phantom and non-sedated pediatric patients: Initial experience with 3rd generation dual-source CT.Z Med Phys. 2016 Dec;26(4):349-361. doi: 10.1016/j.zemedi.2015.11.002. Epub 2015 Dec 15. Z Med Phys. 2016. PMID: 26702762
-
Effect of x-ray tube parameters and iodine concentration on image quality and radiation dose in cerebral pediatric and adult CT angiography: a phantom study.Invest Radiol. 2013 Apr;48(4):192-9. doi: 10.1097/RLI.0b013e31827efc17. Invest Radiol. 2013. PMID: 23344518
-
Automated low-kilovoltage selection in pediatric computed tomography angiography: phantom study evaluating effects on radiation dose and image quality.Invest Radiol. 2013 Aug;48(8):584-9. doi: 10.1097/RLI.0b013e318289f918. Invest Radiol. 2013. PMID: 23563195
-
Pediatric thoracic CT angiography at 70 kV: a phantom study to investigate the effects on image quality and radiation dose.Pediatr Radiol. 2016 Jul;46(8):1114-9. doi: 10.1007/s00247-016-3588-9. Epub 2016 Mar 17. Pediatr Radiol. 2016. PMID: 26987734
Cited by
-
Optimal image quality and radiation doses with optimal tube voltages/currents for pediatric anthropomorphic phantom brains.PLoS One. 2024 Jul 22;19(7):e0306857. doi: 10.1371/journal.pone.0306857. eCollection 2024. PLoS One. 2024. PMID: 39037987 Free PMC article.
-
The changing use of pediatric CT in Australia.Pediatr Radiol. 2016 Jul;46(8):1199-208. doi: 10.1007/s00247-016-3563-5. Epub 2016 Mar 7. Pediatr Radiol. 2016. PMID: 26952101
-
Virtual modeling, stereolithography, and intraoperative CT guidance for the optimization of sagittal synostosis reconstruction: a technical note.Childs Nerv Syst. 2018 May;34(5):965-970. doi: 10.1007/s00381-018-3746-5. Epub 2018 Feb 19. Childs Nerv Syst. 2018. PMID: 29460063
-
Organ-based tube current modulation and bismuth eye shielding in pediatric head computed tomography.Pediatr Radiol. 2022 Dec;52(13):2584-2594. doi: 10.1007/s00247-022-05410-x. Epub 2022 Jul 15. Pediatr Radiol. 2022. PMID: 35836016
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical