. 2016 Oct;30(5):641-7.
doi: 10.1007/s10877-015-9755-y. Epub 2015 Aug 20.

Continuous-wave near-infrared spectroscopy is not related to brain tissue oxygen tension

Affiliations

Continuous-wave near-infrared spectroscopy is not related to brain tissue oxygen tension

Thomas Kerz et al. J Clin Monit Comput. 2016 Oct.

Abstract

Near-infrared spectroscopy (NIRS) has gained acceptance for cerebral monitoring, especially during cardiac surgery, though there are few data showing its validity. We therefore aimed to correlate invasive brain tissue oxygen measurements (PtiO2) with the corresponding NIRS-values (regional oxygen saturation, rSO2). We also studied whether NIRS was able to detect ischemic events, defined as a PtiO2-value of <15 mmHg. Eleven patients were studied with invasive brain tissue oxygen monitoring and continuous-wave NIRS. PtiO2-correlation with corresponding NIRS-values was calculated. We found no correlation between PtiO2- and NIRS-readings. Measurement of rSO2 was no better than flipping a coin in the detection of cerebral ischemia when a commonly agreed ischemic PtiO2 cut-off value of <15 mmHg was chosen. Continuous-wave-NIRS was unable to reliably detect ischemic cerebral episodes, defined as a PtiO2 value <15 mmHg. Displayed NIRS-values did not correlate with invasively measured PtiO2-values. CW-NIRS should not be used for the detection of cerebral ischemia.

Keywords: Brain death; Brain oxygen tension; Brain trauma; Near-infrared spectroscopy; Subarachnoid haemorrhage.

PubMed Disclaimer

References

    1. Stroke. 2014 May;45(5):1269-74 - PubMed
    1. Perioper Med (Lond). 2013 May 19;2(1):11 - PubMed
    1. J Clin Monit Comput. 2015 Apr;29(2):217-20 - PubMed
    1. Neurocrit Care. 2008;9(3):326-31 - PubMed
    1. Neurosurg Focus. 2012 Feb;32(2):E2 - PubMed

MeSH terms