[Augmentation technique on the proximal humerus]

[Article in German]
Affiliations
Randomized Controlled Trial

[Augmentation technique on the proximal humerus]

[Article in German]
A Scola et al. Unfallchirurg. 2015 Sep.

Abstract

Background and objectives: The treatment of osteoporotic fractures is still a challenge. The advantages of augmentation with respect to primary in vitro stability and the clinical use for the proximal humerus are presented in this article.

Material and methods: In this study six paired human humeri were randomized into an augmented and a non-augmented group. Osteosynthesis was performed with a PHILOS plate (Synthes®). In the augmented group the two screws finding purchase in the weakest cancellous bone were augmented. The specimens were tested in a 3-part fracture model in a varus bending test.

Results: The augmented PHILOS plates withstood significantly more load cycles until failure. The correlation to bone mineral density (BMD) showed that augmentation could partially compensate for low BMD.

Conclusion: The augmentation of the screws in locked plating in a proximal humerus fracture model is effective in improving the primary stability in a cyclic varus bending test. The targeted augmentation of two particular screws in a region of low bone quality within the humeral head was almost as effective as four screws with twice the amount of bone cement. Screw augmentation combined with a knowledge of the local bone quality could be more effective in enhancing the primary stability of a proximal humerus locking plate because the effect of augmentation can be exploited more effectively limiting it to the degree required. The technique of augmentation is simple and can be applied in open and minimally invasive procedures. When the correct procedure is used, complications (cement leakage into the joint) can be avoided.

PubMed Disclaimer

References

    1. J Bone Joint Surg Am. 2002 Nov;84-A(11):1919-25 - PubMed
    1. J Bone Joint Surg Am. 2009 Jun;91(6):1320-8 - PubMed
    1. J Orthop Trauma. 2007 Oct;21(9):621-7 - PubMed
    1. Arch Orthop Trauma Surg. 2014 Mar;134(3):343-9 - PubMed
    1. J Trauma Acute Care Surg. 2012 Apr;72(4):1098-101 - PubMed

Publication types

MeSH terms

LinkOut - more resources