Randomized Controlled Trial
. 2015 Dec;26(4):359-66.
doi: 10.1007/s00399-015-0394-2. Epub 2015 Aug 28.

Minimizing right ventricular pacing in pacemaker patients with intact and compromised atrioventricular conduction : Results from the EVITA Trial

Affiliations
Randomized Controlled Trial

Minimizing right ventricular pacing in pacemaker patients with intact and compromised atrioventricular conduction : Results from the EVITA Trial

A Bauer et al. Herzschrittmacherther Elektrophysiol. 2015 Dec.

Abstract

Introduction: Unnecessary ventricular pacing is associated with increased morbidity and mortality. Over the years different algorithms have been developed to reduce right ventricular pacing.

Objectives: Goal of the present study was to test the efficacy of the ventricular intrinsic preference (VIP) algorithm in patients with atrioventricular intact (AVi) and atrioventricular compromised (AVc) AV-conduction.

Methods: Evaluation of VIP feature in pacemaker patients (EVITA) was a multicenter, prospective, randomized trial (Trials.gov Identifier: NCT00366158). In total, 389 patients were randomized to AVc group: n = 140/132 VIP OFF/VIP On, AVi group: n = 54/63 VIP OFF/VIP ON). One-month post-implantation AV conduction testing (AVc: PR/AR interval > 210 ms) was performed. Follow-up visits occurred 6 and 12 months after DDD-pacemaker implantation.

Results: In AVi and AVc-patients initiation of the VIP feature significantly reduced incidence of ventricular pacing (AVi: 53 ± 38 vs. 9 ± 21%, p = 0.0001; AVc: 79 ± 31 vs. 28 ± 35%, p = 0.0001). DDD-pacemaker implantation per se significantly reduced incidence of AF in VIP ON (AVi 27 vs. 0%, p < 0.0001; AVc 29 vs. 3%, p < 0.0001) and VIP OFF patients (AVi 43 vs. 4%, p < 0.0001; AVc 33 vs. 3, p < 0.0001), without significant differences between VIP ON and OFF groups (p > 0.05). In the AVc group activation of VIP significantly reduced incidence of adverse events (AE). All-cause mortality was not significantly different in VIP ON (n = 5) and VIP OFF (n = 4, p > 0.05) patients.

Conclusion: AV search hysteresis (VIP) markedly reduces ventricular pacing both in patients with normal AV conduction and in patients with prolonged PR interval or intermittent AV block.

Keywords: Atrioventricular-hysteresis; Electrophysiology; Heart failure; Pacemaker.

PubMed Disclaimer

References

    1. Circulation. 1998 Sep 29;98(13):1315-21 - PubMed
    1. Pacing Clin Electrophysiol. 1992 Nov;15(11 Pt 2):1927-31 - PubMed
    1. Europace. 2010 Jan;12(1):96-102 - PubMed
    1. Europace. 2011 Jun;13(6):765-7 - PubMed
    1. Pacing Clin Electrophysiol. 2006 Apr;29(4):406-12 - PubMed

Publication types

MeSH terms