Eye lens monitoring for interventional radiology personnel: dosemeters, calibration and practical aspects of H p (3) monitoring. A 2015 review
- PMID: 26343787
- DOI: 10.1088/0952-4746/35/3/R17
Eye lens monitoring for interventional radiology personnel: dosemeters, calibration and practical aspects of H p (3) monitoring. A 2015 review
Abstract
A thorough literature review about the current situation on the implementation of eye lens monitoring has been performed in order to provide recommendations regarding dosemeter types, calibration procedures and practical aspects of eye lens monitoring for interventional radiology personnel. Most relevant data and recommendations from about 100 papers have been analysed and classified in the following topics: challenges of today in eye lens monitoring; conversion coefficients, phantoms and calibration procedures for eye lens dose evaluation; correction factors and dosemeters for eye lens dose measurements; dosemeter position and influence of protective devices. The major findings of the review can be summarised as follows: the recommended operational quantity for the eye lens monitoring is H p (3). At present, several dosemeters are available for eye lens monitoring and calibration procedures are being developed. However, in practice, very often, alternative methods are used to assess the dose to the eye lens. A summary of correction factors found in the literature for the assessment of the eye lens dose is provided. These factors can give an estimation of the eye lens dose when alternative methods, such as the use of a whole body dosemeter, are used. A wide range of values is found, thus indicating the large uncertainty associated with these simplified methods. Reduction factors from most common protective devices obtained experimentally and using Monte Carlo calculations are presented. The paper concludes that the use of a dosemeter placed at collar level outside the lead apron can provide a useful first estimate of the eye lens exposure. However, for workplaces with estimated annual equivalent dose to the eye lens close to the dose limit, specific eye lens monitoring should be performed. Finally, training of the involved medical staff on the risks of ionising radiation for the eye lens and on the correct use of protective systems is strongly recommended.
Similar articles
-
Eye lens dose in interventional cardiology.Radiat Prot Dosimetry. 2015 Jul;165(1-4):289-93. doi: 10.1093/rpd/ncv051. Epub 2015 Mar 24. Radiat Prot Dosimetry. 2015. PMID: 25809107
-
ASSESSMENT OF EYE LENS DOSES IN INTERVENTIONAL RADIOLOGY: A SIMULATION IN LABORATORY CONDITIONS.Radiat Prot Dosimetry. 2016 Sep;170(1-4):256-60. doi: 10.1093/rpd/ncv376. Epub 2015 Jul 28. Radiat Prot Dosimetry. 2016. PMID: 26224737
-
EYE LENS RADIATION EXPOSURE OF WORKERS DURING MEDICAL INTERVENTIONAL PROCEDURES AND SURGERY.Radiat Prot Dosimetry. 2018 Dec 1;182(3):323-328. doi: 10.1093/rpd/ncy068. Radiat Prot Dosimetry. 2018. PMID: 30590843 Clinical Trial.
-
A REVIEW OF TWO METHODS USED IN THE USA TO ASSESS HE DURING FLUOROSCOPIC-BASED RADIOLOGY.Radiat Prot Dosimetry. 2016 Sep;170(1-4):307-10. doi: 10.1093/rpd/ncw006. Epub 2016 Feb 17. Radiat Prot Dosimetry. 2016. PMID: 26891788 Review.
-
Eye dosimetry in interventional radiology and cardiology: current challenges and practical considerations.Radiat Prot Dosimetry. 2014 Dec;162(3):329-37. doi: 10.1093/rpd/nct291. Epub 2013 Nov 20. Radiat Prot Dosimetry. 2014. PMID: 24262928 Review.
Cited by
-
Evaluation of 3D Printed Anthropomorphic Head Phantom for Calibration of TLD in Eye Lens Dosimeter.J Biomed Phys Eng. 2023 Apr 1;13(2):193-202. doi: 10.31661/jbpe.v0i0.2109-1401. eCollection 2023 Apr. J Biomed Phys Eng. 2023. PMID: 37082553 Free PMC article.
-
New organ-based tube current modulation method to reduce the radiation dose during computed tomography of the head: evaluation of image quality and radiation dose to the eyes in the phantom study.Radiol Med. 2017 Aug;122(8):601-608. doi: 10.1007/s11547-017-0755-5. Epub 2017 Mar 24. Radiol Med. 2017. PMID: 28341967
-
Radiation Dose to the Eye Lens Through Radiological Imaging Procedures at the Surgical Workplace During Trauma Surgery.Int J Environ Res Public Health. 2019 Oct 11;16(20):3850. doi: 10.3390/ijerph16203850. Int J Environ Res Public Health. 2019. PMID: 31614696 Free PMC article.
-
Angular dependence of shielding effect of radiation protective eyewear for radiation protection of crystalline lens.Radiol Phys Technol. 2019 Dec;12(4):401-408. doi: 10.1007/s12194-019-00538-2. Epub 2019 Oct 15. Radiol Phys Technol. 2019. PMID: 31617146
-
Summary of the Italian inter-society recommendations for radiation protection optimization in interventional radiology.Radiol Med. 2018 May;123(5):378-384. doi: 10.1007/s11547-017-0849-0. Epub 2018 Jan 6. Radiol Med. 2018. PMID: 29307078
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical